Keysight MXR Series vs. 6 Series B MSO Mixed Signal Oscilloscope Comparison


Oscilloscope
Oscilloscope Performance Specs
Tektronix 6 Series B MSO
✔ 10 GHz max BW & 50 GS/s max SR
✔ 10 GHz & 25 GS/s on four channels
5 GHz & 12.5 GS/s on eight channels
✔ 12-bit Analog-to-Digital Converter (ADC)
✔ Up to 64 digital channels (500MHz, 25GS/s)
✔ 500,000 wfms/s capture rate with Fast Acq
✔ All trigger types available on ALL 8 channels
Industry's Only Std. Closed Embedded OS or optional Windows 10 OS
Keysight MXR
6 GHz max BW and 16 GS/s max SR
6 GHz and 16 GS/s on four channels
✔ 6 GHz and 16 GS/s on eight channels
10-bit Analog-to-Digital Converter
16 digital channels (300 MHz, 8 GS/s)
200,000 wfms/s max capture rate
Pulse width, runt, timeout, burst, more triggers ONLY available on channels 1-4
Windows 10 OS Only
Best in Class Noise Performance1,2
Bandwidth Volts / Div 6 Series B MSO MXR
1 GHz 1 mV 51.1 µV ✔ 73 µV
10 mV 82.9 µV ✔ 99 µV
100 mV 829 µV ✔ 821 µV
1 V 6.22 mV ✔ 6.33 mV
4 GHz 1 mV 97.4 µV ✔ 132 µV
10 mV 171 µV ✔ 189 µV
100 mV 1.73 mV ✔ 1.46 mV
1 V 1.73 mV ✔ 1.46 mV
6 GHz 1 mV 124 µV ✔ 193 µV
10 mV 197 µV ✔ 251 µV
100 mV 2.39 mV ✔ 2.03 mV
1 V 19 mV ✔ 16.26 mV
Vertical Resolution and Signal Clarity
Tektronix 6 Series B MSO
✔ ENOB: 8.2 bits (1 GHz), 7.7 bits (2.5 GHz), 7.2 bits (4 GHz), 6.8 bits (6 GHz)
✔ Signal-to-Noise Ratio: -36 dB (1 mV/div), -51 dB (10 mV/div), -53 dB (1 V/div)
Keysight MXR
ENOB: 8.0 bits (1 GHz), 7.6 bits (2.5 GHz), 7.2 bits (4 GHz), 6.8 bits (6 GHz)
Signal-to-Noise Ratio: -32 dB (1 mV/div), -49 dB (10 mV/div), -51 dB (1 V/div)
Spectrum Analysis
Tektronix 6 Series B MSO
✔ Spectrum View w/ 2 GHz span FREE
    0-10 GHz center frequency FREE
✔ Simultaneous view of time and frequency domains
✔ RF vs. Time waveforms (mag, freq, phase) and RF vs. Time trigger optional
Keysight MXR
RTSA with 2 GHz span PAID OPTION
    0-6 GHz center frequency PAID OPTION
Time and frequency analysis are separate operating modes; no correlation on same input
No RF vs. Time analysis or triggering capabilities
key features comparison key features comparison

** ENOB was tested at 500mV Full Scale and 90% of screen