I use a TDS5034B with the TDSPWR3 power analysis software to perform SOA test. I want to set the record length to maximum so that I can capture the waveform a with high sample rate. But when I perform SOA testing+ the scope will display an error window that I should set the record length to less than 200K data points.

質問:

I use a TDS5034B with the TDSPWR3 power analysis software to perform SOA test. I want to set the record length to maximum so that I can capture the waveform a with high sample rate. But when I perform SOA testing+ the scope will display an error window that I should set the record length to less than 200K data points.

回答:

Please update the TDSPWR3 software to the latest version. This is available for download from the Tektronix web site, under Software and Drivers.