On the AWG2000 series+ if I select the FG mode+ what is the maximum frequency that can be selected?

質問:

On the AWG2000 series+ if I select the FG mode+ what is the maximum frequency that can be selected?

回答:

In FG mode the AWG is outputting pre-defined waveform data.  To maintain a minimum level of waveform quality there is a limit of 100 points per cycle, depending on the model.  This means that the maximum clock rate of the instrument would be divided by 100 to determine the maximum frequency of a function generator mode waveform.  See table below.
 

  Model      

Max Clock     

Max FG freq.

AWG2005

  20 MHz

  200 KHz

AWG2020

 250 MHz

   2.5 MHz

AWG2021

 250 MHz

   2.5 MHz

AWG2040

     1 GHz

    10 MHz

AWG2041

     1 GHz

    10 MHz

 

 