Did the Performance Verification software change when TLA Application Software v5.1 was introduced?

Question :

Did the Performance Verification software change when TLA Application Software v5.1 was introduced?

Answer :

The Performance Verification software did change a little.  The software for TLA7Lx/Mx/Nx/Px/Qx and TLA7Dx/Ex modules have not been updated to run on a remote PC controller.  If you need to verify one of those modules, you must do so in a TLA7012 or TLA700-series mainframe.  Also, the PV software will display a PCI-VXI bridge error when running with TLA Application Software version 5.x, unless you first delete the Visa32.dll file from the C:\Tekcats folder.  Then when you see the message "UNABLE TO ENABLE SRQ EVENTS", you can continue by pressing the Enter key twice.

This FAQ Applies to:

Product Series: TLA7000 TLA7000 TLA7000

Product:

FAQ ID 55391