    2450 - Range Compliance: Source limited unexpectedly due to low Measurement Range
    Question: When we try to read the measurements from our 2450 instrument it causes any output voltage to automatically be 1.154 volts into a DMM, which is below the programmed value. Answer:  There is a condition on 24xx series SMUs called ‘range …
    KB-01028
    Sink function doesn't go to zero when not initialized
    Question: I've set my 2450 to sink 100mA. When I enable the load it sinks 100mA, but when I disable the load with blue button in the corner, it still sinks 10mA when the output if off. I want it to go to zero mA load. Answer: The 2450 SMU is a four …
    KB-01047
    Sweep: Can I use an external trigger to stop a SourceMeter sweep?
    Question: Sweep: Can I use an external trigger to stop a SourceMeter sweep? Answer: The SourceMeter trigger model does not have provisions for stopping a sweep using a trigger signal. However, there is a way to accomplish this by using the …
    KB-01468
    Speed: What is NPLC?
    Question: Speed: What is NPLC?   Answer: The Number of Power Line Cycles (NPLC) indicates how long an input signal is integrated to obtain a single measurement. Noise introduced from the power line tends to be periodic. If the A/D converter …
    KB-01485
    Remote Operation: Why do I get a VISA timeout (BFFF0015) using RS-232 with a SourceMeter?
    Question: Why do I get a VISA timeout (BFFF0015) using RS-232 with a SourceMeter?   Answer: When programming a SourceMeter through the VISA layer into RS-232, the use of SCPI commands is required. However, there are additional subtle requirements …
    KB-01460
    Ohms: What is AUTO or MANUAL mode when measuring Resistance with a 2400?
    Question: Ohms: What is AUTO or MANUAL mode when measuring Resistance with a 2400? Answer: What is the difference between Auto and Manual mode for resistance in the Model 2400? The AUTO resistance measuring mode of the Model 2400 configures the unit …
    KB-01492
    How do I create an I-V Characterizer test that uses more than one SMU Instrument?
    Question: How do I create an I-V Characterizer test that uses more than one SMU Instrument? Answer: Use the “Instruments” button in the I-V Characterizer app toolbar to access the “Add Channels” button to add more SMU instruments to your test.   You …
    KB-01364