Question :
How does the support offered by the DPO4USB application module for the MSO4000 and DPO4000 Series oscilloscopes vary for the different USB speeds?
Answer :
The DPO4USB application module provides the following support for the different USB speeds:
Low-speed (1.5 Mb/s) USB triggering, decode, and search features include:
- Single-ended and differential input signals
- ASCII, hexadecimal, decimal, and binary display formats
- Trigger on: Reset, Suspend, Resume, End of Packet, Sync, Token (address and end point), Data Packet, Handshake Packet, Special Packet, Error
- Bus Decode with Event table and save to .CSV
- Automatic Search: same criteria as triggering
- Supported on MSO4104, MSO4054, MSO4034, MSO4032, DPO4104, DPO4054, DPO4034, DPO4032
Full-speed (12 Mb/s) USB triggering, decode, and search features include:
- Single-ended and differential input signal formats:
- ASCII, hexadecimal, decimal, and binary display formats
- Trigger on: Reset, Suspend, Resume, End of Packet, Sync, Token (address and end point), Data Packet, Handshake Packet, Special Packet, Error
- Bus Decode with Event table and save to .CSV
- Automatic Search: same criteria as triggering
- Supported on MSO4104, MSO4054, MSO4034, MSO4032, DPO4104, DPO4054, DPO4034, DPO4032
High-speed (480 Mb/s) USB decode and search features include:
- Differential (only) input signals
- ASCII, hexadecimal, decimal, and binary display formats
- Edge Trigger (only)
- Bus Decode with Event table and save to .CSV
- Automatic Search: same criteria as low-speed and full-speed triggering
- Supported on MSO4104 and DPO4104
This FAQ Applies to:
No product series
Product:
FAQ ID 59491