How does the support offered by the DPO4USB application module for the MSO4000 and DPO4000 Series oscilloscopes vary for the different USB speeds?

Question :

How does the support offered by the DPO4USB application module for the MSO4000 and DPO4000 Series oscilloscopes vary for the different USB speeds?

Réponses :

The DPO4USB application module provides the following support for the different USB speeds:

Low-speed (1.5 Mb/s) USB triggering, decode, and search features include:

  • Single-ended and differential input signals
  • ASCII, hexadecimal, decimal, and binary display formats
  • Trigger on: Reset, Suspend, Resume, End of Packet, Sync, Token (address and end point), Data Packet, Handshake Packet, Special Packet, Error
  • Bus Decode with Event table and save to .CSV
  • Automatic Search: same criteria as triggering
  • Supported on MSO4104, MSO4054, MSO4034, MSO4032, DPO4104, DPO4054, DPO4034, DPO4032

Full-speed (12 Mb/s) USB triggering, decode, and search features include:

  • Single-ended and differential input signal formats:
  • ASCII, hexadecimal, decimal, and binary display formats
  • Trigger on: Reset, Suspend, Resume, End of Packet, Sync, Token (address and end point), Data Packet, Handshake Packet, Special Packet, Error
  • Bus Decode with Event table and save to .CSV
  • Automatic Search: same criteria as triggering
  • Supported on MSO4104, MSO4054, MSO4034, MSO4032, DPO4104, DPO4054, DPO4034, DPO4032

High-speed (480 Mb/s) USB decode and search features include:

  • Differential (only) input signals
  • ASCII, hexadecimal, decimal, and binary display formats
  • Edge Trigger (only)
  • Bus Decode with Event table and save to .CSV
  • Automatic Search: same criteria as low-speed and full-speed triggering
  • Supported on MSO4104 and DPO4104