In the TLA Application Software Listing Window I would like to know the State for each acquired sample (Trigger State Machine).

Question:

In the TLA Application Software Listing Window I would like to know the State for each acquired sample (Trigger State Machine).

Answer:

Go the the Setup window and add a new Group. Then add the status bits to this group. They are located at the bottom of the section called "Probe Channels/Names". Now right click the new Group Name for your status bits and choose "Select Status Bits Type". You can choose between Trigger State Machine (TSM), Clocking State Maching (CSM) or Clocking Group Complete (CGC).

This FAQ Applies to:

Product Series: TLA7000, TLA7000, TLA7000

Product: TLA704, TLA711, TLA714, TLA721, TLA601, TLA602, TLA603, TLA604, TLA611, TLA612, TLA613, TLA614, TLA621, TLA622, TLA623, TLA624, TLA7Q4, TLA7AB4, TLA7N1, TLA7N2, TLA7N3, TLA7N4, TLA7P2, TLA7P4, TLA7Q2, TLA7AA1, TLA7AA2, TLA7AA3, TLA7AA4, TLA7AB2, TLA7L1, TLA7L2, TLA7L3, TLA7L4, TLA7M1, TLA7M2, TLA7M3, TLA7M4, TLA5201, TLA5202, TLA5203, TLA5204, TLA7NA1, TLA7NA2, TLA7NA3, TLA7NA4, TLA7012, TLA7016

FAQ ID : 55856