Contact us

Live Chat with Tek representatives. Available 6:00 AM - 4:30 PM

Call

Call us at

Available 6:00 AM – 5:00 PM (PST) Business Days

Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

I am using the 80SSPAR software and the Iconnect S-parameter Wizard software and Some of the waveforms are not positioned properly on screen.

Question :

I am using the 80SSPAR software and the Iconnect S-parameter Wizard software and Some of the waveforms are not positioned properly on screen.

Answer :

The deskew process, before measuring the first DUT, had trouble.  If using a test fixture to connect the DUT, perform the deskew with just the cables.


This FAQ Applies to:

No product series

Product:

FAQ ID 59531

View all FAQs »