  • 常見問答集 常見問答集 ID
    How many multiplexer cards can the DAQ6510 hold?
    Question: How many multiplexer cards can the DAQ6510 hold? Answer: There are two slots for plug in modules. The DAQ6510 can accept the various 12 switching cards available, but it only hold 2 at a time. If you have higher density switching, it is …
    KB-01197
    DAQ6510 Data in Incorrect Columns in export or KickStart
    Problem: Sometimes data from a DAQ6510, whether exported to CSV or to KickStart ends up in the wrong column or area. This is an issue with commas appearing in the wrong place causing the data to be allocated to the wrong location.   Solution: If this …
    KB-04161
    DAQ/DMM 65XX/7510 startup errors
    Question: Some units on startup report an error "Command requires installation of accessory card" (1151) as well as an error claiming that a script or line of code cannot be executed. The instrument may also then not respond to some commands or …
    KB-01148
    DAQ6510 -285 TSP Syntax error on line 1113 upon device startup
    Question: Some DAQ6510 units are encountering an issue upon startup where they report a -285 "TSP Syntax error at line 1113: unfinished string near '' " error. This does not seem to impact device functionality but is concerning and irritating to …
    KB-01623
    Accessory for 20 Channels Data Acquisition from DAQ 6510
    Question: How to have 20 channels data acquisition from the DAQ6510, what accessory or module is needed? Answer: To achieve 20 channels data acquisition with the DAQ6510, you have several optional cards to choose from. The most common and recommended …