TDSHT3 shows that the signal overshoot and undershoot is high. What can I do?

Question :

TDSHT3 shows that the signal overshoot and undershoot is high. What can I do?

Réponses :

If you suspect the probing is causing the Overshoot and undershoot problem, please consider the following:
  • Make sure that you are using the correct probe. Check the signal rise time and Probe’s typical rise time. If the probe typical rise time is greater than the signal rise time, the overshoot is expected. Use a higher end probe for the measurement.
  • Try avoiding using probe adaptors. (Square pin adaptors may introduce overshoots)