Is BUS trigger mode supported in the AB sequence trigger mode on the 5 series MSO?

Subject: Setup & Operation | Published: October 2026 | FAQ ID: KB-01349

Question: Is BUS trigger mode supported in the AB sequence trigger mode on the 5 series MSO?

Answer: Yes. However, it does not apply to both events. It could be only either A or B event for the bus trigger. The second event is locked to edge trigger type.