Question :
What is the Input Resistance of the 6487 when Zero Check is enabled?
Réponses :
Product(s) :Picoammeter: 6487
Question : What is the Input Resistance of the 6487 when Zero Check is enabled?
Answer: A model 6517A was used to apply a 1 volt excitation to the 6487. The following values of resistance were measured:
RANGE INPUT RESISTANCE 20mA or 2mA 510 ohms 200uA or 20uA 40 kohm 2uA or 200nA 3.5 Mohm 20nA or 2nA 12 Mohm
Numéro de la FAQ 71476