Question :
Are Active Probes with the Tek Probe Interface+ such as the P6330+ compatible with the 80A03?
Réponses :
No. The 80A03 only recognizes the probe with the TekConnect interface (used by the P7000 series probes). The 80A03 does not recognize the TCA-BNC so it will not provide power probes such as the P6330.
In addition, the TCA-SMA, TCA-N, and TCA-1Meg adaptors are not recognized by the 80A03.