問題:
When using pulse width trigger on the DPO4000/MSO4000+ why does the trigger indicator not line up with the start of the pulse?
答案:
The trigger lines up with the end of the pulse. The oscilloscope must first find a pulse that corresponds to the trigger conditions set, and at that point triggers the scope.
此常見問答集適用於:
沒有任何產品系列
產品:
常見問答集 ID 56196
檢視所有常見問答集 »