질문:
Can the BNC output shield (outside of the BNC connector) of the AFG310 and AFG320 be elevated above ground?
답변:
Yes, the shield on the BNC connector is not ground referenced and can be elevated as high as 42V peak (AC + DC) or 30 Vrms above earth ground.
이 FAQ는 다음에 적용 됩니다:
제품 시리즈 없음
제품:
FAQ ID 57191