Why do I get a flat line signal in Vectorscope Display Mode when I apply a 75% or 100% Color Bar Composite Video signal to Channel 1 of my TDS3000 series scope with TDS3VID or TDS3SDI Video module?
答案:
The TDS3VID, or TDS3SDI Vectorscope feature only supports Component Video signals. You must apply a Color Difference Component Analog Video signal, for example; Y,Pb,Pr to two input channels on the TDS3000/B Scope which uses XY mode to produce the Vectorscope display.