與我們聯絡

與 Tek 業務代表即時對談。 上班時間:上午 6:00 - 下午 4:30 (太平洋時間)

致電

請致電

與 Tek 業務代表即時對談。 上班時間:上午 8:30 - 下午 5:30 (太平洋時間)

下載

下載手冊、產品規格表、軟體等等:

下載類型
機型或關鍵字

意見回饋

I have connected my OTDR to my fiber under test and when a test is started I receive error message "Poor Front panel connection". What are the possible causes and solutions to this error?

問題:

I have connected my OTDR to my fiber under test and when a test is started I receive error message "Poor Front panel connection". What are the possible causes and solutions to this error?

答案:

1. The connection at the front panel is dirty.

Clean fiber

2. Fiber end on OTDR is damaged.

Send OTDR in for repair

3. The launch cable is bad.

Replace cable

4. The fiber under test is to short for the test mode selected.

When InteliTrace, or End of Fiber setting is selected you need at least 100+ meters of fiber connected.

If the fiber is short use Manual setting and use a short pulse. If you trust that the connection is good then continue the test.

5. Laser/detector in OTDR is or has failed.

Send in for repair.


此常見問答集適用於:

沒有任何產品系列

產品:

常見問答集 ID 59561

檢視所有常見問答集 »