Is the GPIB command set the same for the TDS1000+TDS2000+ TDS1000B+ and TDS2000B Series as the TDS200 Series?

問題:

Is the GPIB command set the same for the TDS1000+TDS2000+ TDS1000B+ and TDS2000B Series as the TDS200 Series?

答案:

There are minor differences to support new features. See below.

Programming language differences between TDS200 and TDS1000/2000 Series
(Last revision: 10-Feb-2003)
This is just a summary of what changed. See Programmer Manual for details of syntax, new arguments, etc.

Command or query Change
autoset:signal? New header subfunction
autoset:view New header subfunction
display:invert New header subfunction
hardcopy:format More argument choices
hardcopy:inksaver New header subfunction
horizontal:main:scale Increased range
measurement:immed:type More argument choices
measurement:meas Maximum increased from 4 to 5
measurement:meas:type More argument choices
recall:setup Maximum increased from 5 to 10
save:setup Maximum increased from 5 to 10
trigger:main:edge:source New argument choices for 4-channel scopes
trigger:main:frequency? New header subfunction
trigger:main:pulse New header subfunction
trigger:main:pulse:source New header subfunction
trigger:main:pulse:width New header subfunction
trigger:main:pulse:width:polarity New header subfunction
trigger:main:pulse:width:when New header subfunction
trigger:main:pulse:width:width New header subfunction
trigger:main:type New argument (pulse)
trigger:main:video:line New header subfunction
trigger:main:video:standard New header subfunction
trigger:main:video:sync More argument choices