How do you calculate the amount of time the Spectrum Time indicator (the orange bar) represents?
Spectrum Time = Window Factor / Resolution BandwidthWindow Factors: FFT WindowFactor Kaiser2.23 Rectangular0.89 Hamming1.30 Hanning1.44 Blackman-Harris1.90 Flat-Top3.77 The Resolution Bandwidth can be found next to the RBW readout on the MDO4000 display.

This FAQ Applies to:
50016
After sending the CURVE? query why do I only get 5000 points even though the record length is more than 5000 points?
It is possible the DATA:STOP value is set to 5000 which is a common default value. When transferring data from the oscilloscope the user must specify a DATA:START and DATA:STOP value or the default values will be used.

This FAQ Applies to:
52346
Are there IBIS or Spice loading models available for your Logic Analyzer probes?
Unfortunately we do not have pre-developed IBIS/Spice Models using the information provided in the P68xx probe manual.

This FAQ Applies to:
52351
Are there programming examples of using TekVisa?
When TekVisa is installed a folder named “Visa Samples” is also installed in the C:\VXIpnp\WINNT\TekVISA\VisaSamples folder. There are also examples in the Tekvisa programmers manual located at: /dpo3012-manual/tekvisa An alternative to this sample …

This FAQ Applies to:
52361
AWG5000: How do I enable the second monitor?
Minimize the AWG application then right click on an empty area of the screen. Select Properties then the Settings tab and click on monitor 2. Now check “extend my Windows desktop to this monitor” and press Apply.

This FAQ Applies to:
52366
AWG5000: What is the SFDR on the AWG5000?
SFDR for the AWG5k series is shown in the list below.AWG5012 and AWG5014Normal Mode:50 dBc (37.5 MHz, 1.2 GS/s, 2.0 Vpk-pk)60 dBc (10 MHz, 600 MS/s, 1.0 Vpk-pk)80 dBc (1 MHz, 600 MS/s, 1.0 Vpk-pkDirect Mode:64 dBc (10 MHz, 600 MS/s, 0.6 Vpk-pk)80 dBc (1 …

This FAQ Applies to:
52371
AWG7000: How do I enable the second monitor?
Minimize the AWG application then right click on an empty area of the screen. Select Properties then the Settings tab and click on monitor 2. Now check “extend my Windows desktop to this monitor” and press Apply.

This FAQ Applies to:
52376
Can a Tektronix Encore product be reserved?
Reconditioned products are subject to availability upon receipt of a purchase order. A unit cannot be reserved without a purchase order.

This FAQ Applies to:
52381
Can I create a file (reference memory) in an external application and have the Logic Analyzer use the compare function with acquired data to compare the differences between the two?
The TLA Application Software cannot compare against an external source. The target has to be another saved .tla file acquired through the TLA probe interface.

This FAQ Applies to:
52386
Can my TLA600 be upgraded from Windows 2000 to Windows XP and still run version 4.4 of the TLA software?
The TLA600 is not supported under Windows XP. Windows 2000 and application software v4.4 was the last supported software configuration.

This FAQ Applies to:
52391