연락처

텍트로닉스 담당자와 실시간 상담 6:00am-4:30pm PST에 이용 가능

전화

전화 문의

9:00am-6:00PM KST에 이용 가능

다운로드

매뉴얼, 데이터 시트, 소프트웨어 등을 다운로드할 수 있습니다.

다운로드 유형
모델 또는 키워드

피드백

What is the difference between VXIbus Mainframe total+ usable power?

질문:

What is the difference between VXIbus Mainframe total+ usable power?

답변:

Total power is the maximum power that the power supply can provide. Total power is the sum of V * I for each power rail. VXI power supplies provide 7 voltage rails (24, 12, 5, -2, -5.2, -12, -24). If a power supply only supplied one voltage rail, then total power and usable power would be the same. Since VXI power supplies have 7 rails, they typically cannot drive all 7 rails at maximum current at the same time. Usable power is the most wattage that the power supply can provide at one time. Since usable power is really the maximum output power, total power is relatively meaningless. Usable power specifications are the primary concern of most VXIbus test system integrators.


이 FAQ는 다음에 적용 됩니다:

제품 시리즈 없음

제품:

FAQ ID 62241

모든 FAQ 보기 »