질문:
Can the digital output pulse be longer in duration than the channel time with ExceLINX-1A?
답변:
Yes, but the pulse will never be reset. It will always be HI or LO depending on the configuration. It is best to select an output pulse duration that is shorter in duration than the channel time. For example if the output pulse duration is set for 100msec. Then the channel time must be set for longer than 100msec.
이 FAQ는 다음에 적용 됩니다:
제품 시리즈 없음
제품:
FAQ ID 70086