질문:
Why is the TDS1000/TDS2000 bandwidth limited? The sample rate is 2 .0 GS/s+ Considering Nyquest the scope should have a 1 GHz bandwidth.
답변:
1) The scope over samples, all signals are not sine waves. 2) The scope has an analog front end which is bandwidth limited to 200 MHz.