State VS. Timing - What is a state timing measurement or a timing measurement? What is the difference between internal and external clocking?

질문:

State VS. Timing - What is a state timing measurement or a timing measurement? What is the difference between internal and external clocking?

답변:

For a state measurement the logic analyzer is clocked by a signal from the system under test. At every valid clock, the logic analyzer takes a snapshot of the system under test. This snapshot (sample) is clocked by the system so the measurements are synchronous with the system under test.

For a timing measurement, the logic analyzer samples data at regular intervals (500ps*-50ms) chosen in the clocking setup. This internal clock is not related to the system under test so the measurements are asynchronous to the system under test.

*See TLA module specifications for the minimum Deep Memory Timing


이 FAQ는 다음에 적용 됩니다:

제품 시리즈: TLA7000