질문:
My ISA Board Works When I Take Just a Single Sample, but Does Not When Sampling at a Rate
답변:
If the board returns correct data when taking single samples but fails to operate satisfactorily when acquiring data at a programmed rate, there is probaby an interrupt or DMA channel conflict. The single sample mode does not require the use of an interrupt or DMA transfer. When a pacer clock is used to control the rate at which data are acquired, the IRQ and/or DMA are necessary to coordinate these transfers. Sound cards often present DMA conflicts. Network or even video cards often conflict with interrupt settings.
이 FAQ는 다음에 적용 됩니다:
제품 시리즈 없음
제품:
FAQ ID 70806
모든 FAQ 보기 »