Häufig gestellte Fragen

FAQ
Where can I find Equivalent circuit models for P7708, P7713, P7716, P7720?
Equivalent circuit models can be found in the P7708, P7713, P7716, P7720 Users manual.  You can download manuals from Tek.com by going to the "Downloads" icon on the right side of the web page.  Select the download type as "manuals", enter in your …

Diese FAQs beziehen sich auf:
776501
When I connect my TBS2000 to the Ethernet via the VXI Web Interface and want to use the control panel it says, “Please enter the password”.  What is the password?
By default, there is no password set, so you can leave the password field blank.   However, a user can set a password through the “Set Password” tab. If a password is set, make sure to note what you set the password to.

Diese FAQs beziehen sich auf:
776436
What happens in High Capacitance mode in Model 2450 SMU?
The SMU is a closed loop control device to maintain a steady force value on your DUT. The capacitance of your cabling and DUT and the current sense resistor (internal to the SMU) form a pole. For certain combinations of current sense resistor and external …

Diese FAQs beziehen sich auf:
775821
What is Source Readback?
You can set the instrument to record and display the voltage or current of the configured source value or the actual source value. When you use the configured source value, the instrument records and displays the value that was configured. When you use the …

Diese FAQs beziehen sich auf:
775816
I trying to install my TekBench AMS license but when I select "Activate" it says "Activation code invalid".
If you are installing an AMS license, use the "Install" button to install the AMS license.  The "Activate" button is only for use internally by Tektronix and is not used when installing AMS licenses.   

Diese FAQs beziehen sich auf:
775586
How do I find the Host ID (HID) in the AFG31000
First the AFG31000 should have code version 1.4.6 or above. Home button, Utility => License Note: Install will be grayed out until a USB thumb drive is in the AFG31000.

Diese FAQs beziehen sich auf:
774761
How to take a screen shot of display of DAQ6510 into USB drive.
To take a screenshot of the graphical waveform on the DAQ6510 all you need to do is press “home” and “enter” at the same time with a thumb drive plugged into the front panel. It will take a screenshot and save it to the USB stick.

Diese FAQs beziehen sich auf:
774756
Does KickStart collect user data?
KickStart optionally collects data about how it is used and transmits this data to a third-party analytics service, which provides the data to Keithley/Tektronix.  KickStart does not collect or transmit personally-identifiable information.  All data …

Diese FAQs beziehen sich auf:
774301
How can I measure current or voltage via LabView for Model 6517B?
This LabView VI allows you to choose: Function (Voltage or Current), NPLC, Counts, Range, Resolution Digits and Buffer Delay Time. The Program outputs the reading buffer data, standard deviation and RMS value. The data will also be saved to a file. This is …

Diese FAQs beziehen sich auf:
774036
What is the effective input resistance on the 6517B electrometer in current mode?
Voltage burden is what is important here. Model 6517B is as close to an ideal ammeter as its voltage burden is extremely low. However, you can calculate the "effective" input resistance of each range. 20pA range is 20uV / 20pA = 1M ohm 200pA range is 100uV …

Diese FAQs beziehen sich auf:
774026