How do I get the Tektronix DPO4000/MSO4000 Oscilloscopes to do iView with the TLA?

Subject: Setup & Operation | Published: October 2026 | FAQ ID: KB-03917

Question: How do I get the Tektronix DPO4000/MSO4000 Oscilloscopes to do iView with the TLA?

Answer: The DPO4000 is unique in that it needs the TEK-USB-488 along with the iView cable to make it work. To set this up you would connect the TEK-USB-488 to the DPO using the USB cable, then connect the GPIB end of the iView cable to the TEK-USB-488 and the USB end into the TLA.

Product Series: TLA7000