Загрузить

Загрузить руководства, технические описания, программное обеспечение и т. д.:

ТИП ЗАГРУЗКИ
МОДЕЛЬ ИЛИ КЛЮЧЕВОЕ СЛОВО

Обратная связь

Analog Oscilloscope

2205

Tektronix больше не продаёт продукты из этого технического описания.
Посмотреть бывшее в употреблении оборудование для тестирования Tektronix Encore.
Просмотреть информацию о технической поддержке и гарантии для этих продуктов.

Features & Benefits

  • 40-MHz Bandwidth Option 40
  • Peak-to-Peak Auto Triggering
  • 0.3-Div Trigger Sensitivity
  • Front-Panel Z-Axis Input
  • 10-ns/div Max. Sweep Rate

Characteristics

Vertical System (2 Identical Channels)

Bandwidth (-3 dB) and Rise Time - 20 MHz and 17.5 ns (5 to 35 °C). 15 MHz and 23.3 ns (0 to 40 °C).

2205 Option 40 Bandwidth (-3 dB) and Rise Time - 40 MHz and 8.75 ns (+5 to +35 °C). 35 MHz and 10 ns (0 to 40 °C).

Deflection Factor and Accuracy - 5 mV/div to 5 V/div ±3%, (15 to 35 °C); ±5%, (0 to 40 °C).

Vertical Operating Modes - CH 1, CH 2, CH 2 IN-VERT, ADD, ALT, CHOP.

CMRR - At least 10:1 at 10 MHz.

Input R and C - 1 Megohm, 25 pF.

Max Input Voltage - 400 V (DC + peak AC), 800 Vpk-pk.

Channel Isolation - 100:1 at 20 MHz.

Horizontal System

Sweep Speeds - 0.5 s/div to 0.1 µs/div (X10 MAG to 10 ns/div)

Accuracy - ±3%; Magnified ±4% (degrade by 1% outside 15 to 35 °C)

Horizontal Operating Modes - X1, X10, X-Y

Trigger System

Trigger Sensitivity -

Internal: 0.30 div at 5 MHz, 1.0 div at 30 MHz

External: 40 mV at 5 MHz, 150 mV at 30 MHz (150 mV at 45 MHz for option 40)

Trigger Operating Modes - Peak-Peak AUTO, NORM, TV FIELD, TV LINE, SGL SWP

Trigger Source - VERT MODE, CH 1, CH 2, LINE, EXT, EXT/10

Trigger Coupling - AC, DC

X-Y Operation

Deflection Factors - Same as vertical system

Bandwidth -

X-Axis: 2 MHz

Y-Axis: Same as vertical system

Phase Difference - ±3 degree from DC to 50 kHz

CRT System

Display - 8 cm × 10 cm, 1.8 kV nominal voltage (2205); 6 kV nominal voltage (Opt. 40).

Controls - INTENSITY, TRACE ROTATION, BEAM FIND, FOCUS.

Z-Axis - 5 V causes noticeable modulation. Usable to 5 MHz.

Power Requirements

Line Voltage Range -

Low: 95 V AC to 128 V AC

High: 185 V AC to 250 V AC

Line Frequency - 48 Hz to 440 Hz

Maximum Power Consumption - 40 W (60 VA)

Environmental Specifications

Instruments meet in part, the environmental requirements of MIL-T-28800D or C for Type III, Class 3, Style D or C equipment as described below.

Ambient Temperature -

Operating: 0 to +40 °C.

Nonoperating: -55 to +75 °C.

Humidity -

Percent: 95% five cycles, (120 hours) 90% +40 °C 4 hours.

Reference: MIL-T-28800D, paragraph 4.5.5.1.2.2.

Tektronix 062-2847-00, ll.

Altitude -

Operating: To 4,500 m (15,000 ft.), maximum operating temp decreases 1 °C per 300 m above 1500 m.

Nonoperating: To 15,000 m (50,000 ft.).

EMC - Meets Class B requirements per VDE 0871-B for radiated and conducted emissions and FCC requirements.

Vibration -

Operating: 15 minutes along each of the 3 major axes, 10 Hz to 55 Hz to 10 Hz in one-minute cycles. Hold for 10 minutes at 55 Hz. Displ. (in pk-pk): 0.015.

Shock -

Operating: 30 g, 1/2 sine, 11-ms duration, 3 shocks per axis along each major axis. Total of 18 shocks.

Other Characteristics

Safety - UL 1244 listed, Certified to CSA C22.2 No. 231-M89

Warranty - 1 year; 3 years optional

Physical Characteristics

Dimensions

mm

in.

Height

137

5.4

Width (with handle)

380

15.0

Depth (without front cover)

440

17.3

Weight

kg

lbs.

Net

6.7

14.8