與我們聯絡

與 Tek 業務代表即時對談。 上班時間:上午 6:00 - 下午 4:30 (太平洋時間)

致電

請致電

與 Tek 業務代表即時對談。 上班時間:上午 8:30 - 下午 5:30 (太平洋時間)

下載

下載手冊、產品規格表、軟體等等:

下載類型
機型或關鍵字

意見回饋

How do I test a SCR for Peak Reverse Blocking Current on my curve tracer?

問題:

How do I test a SCR for Peak Reverse Blocking Current on my curve tracer?

答案:

Peak Reverse Blocking Current - IRRM

 What It Is:

With no gate signal applied, peak reverse blocking current is the maximum reverse current that flows with the SCR in the blocked (Off) state.  Peak reverse blocking current is equal to peak forward blocking current but in the opposite direction.

On the curve tracer, the anode is connected to the Collector Supply and the gate is held open.

What The Display Shows:

The display shows anode voltage on the horizontal axis, and the resulting anode current on the vertical axis.  The specification is met when at the specified reverse voltage, reverse current is less than or equal to the specified maximum.

How To Do It:

1. Set controls:

            A: Max Peak Volts to the lowest setting above the specified VRRM

            B: Max Peak Power Watts to the lowest setting that satisfies (IRRM x VRRM)

            C: Horizontal Volts/Div to display VRRM between the 5th and 10th horizontal divisions

            D: Vertical Current/Div to display IRRM between the 5th and 10th vertical  divisions

            E: Collector Supply Polarity to -DC

            F: Configuration to (Base/Open, Emitter/Common)

            K: Variable Collector Supply to minimum % (full ccw)

            L: Dot Cursor (On)

2. Apply power to the SCR:

            A: Position the Left/Right switch as appropriate

            B: Slowly increase Variable Collector % until the specified VRRM is attained

3. Compare to data sheet specifications:

            A: Check that at the specified VRRM, IRRM is less than or equal to the specified maximum


此常見問答集適用於:

沒有任何產品系列

產品:

常見問答集 ID 52711

檢視所有常見問答集 »