Digital Storage Oscilloscopes

TPS2012B • TPS2014B • TPS2024B

Features & Benefits

  • 100 MHz and 200 MHz Bandwidths
  • Sample Rates up to 2 GS/s Real Time
  • 2 or 4 Fully Isolated and Floating Channels, plus Isolated External Trigger
  • 8 Hours of Continuous Battery Operation with Two Batteries Installed, Hot Swappable for Virtually Unlimited Freedom from AC Line Power
  • Optional Power Application Software offers the Broadest Range of Power Measurements at its Price Point
  • Quickly Document and Analyze Measurement Results with OpenChoice® Software or Integrated CompactFlash® Mass Storage
  • FFT Standard on All Models
  • Advanced Triggers to Quickly Capture the Event of Interest
  • Traditional, Analog-style Knobs and Multilanguage User Interface for Easy Operation
  • Quick Setup and Operation with Autoset Menu, Autorange, Waveform and Setup Memories, and Built-in, Context-sensitive Help
  • Backlit Menu Buttons for High Visibility
  • 11 of the Most Critical Automatic Waveform Measurements

Applications

  • Industrial Power Design, Troubleshooting, Installation, and Maintenance
  • Advanced Electronics Design, Troubleshooting, Installation, and Maintenance
  • Automotive Design and Test
  • Education

TPS2000B Series Oscilloscopes for Powerful Productivity from Bench to Field

The TPS2000B Series offers a distinctive range of capabilities in an oscilloscope with controls and menus you will find familiar and easy to use. Available in 2- or 4-channel versions, the TPS2000B Series with IsolatedChannel™ technology provides isolation from ground and isolation between channels allowing you to take measurements with less worry about damaging circuitry. Battery power comes standard, making it a natural choice for field applications. For work on power electronics, optional software integrates commonly needed power system measurements into the instrument, speeding up power analysis and troubleshooting.

Make Floating and Differential Measurements – Quickly, Accurately, Affordably

Unintentionally grounding a circuit under test is a common cause of poor measurement results and circuit damage. Connecting two or more grounded probes can cause ground loops, and if the current is high enough can result in ruined components and equipment. Most importantly, taking floating measurements without the proper instruments and probes can pose a safety hazard.


Four IsolatedChannel™ inputs and isolated external trigger input for quick, accurate, affordable floating and differential measurements.

Tektronix IsolatedChannel technology simplifies floating measurements. Unlike ground-referenced oscilloscopes, the TPS2000B input connector shells are isolated from each other and from earth ground. Within the specified 600 VRMS maximum float voltage, IsolatedChannel technology keeps current from flowing between the TPS2000B input BNC shells or from any BNC shell to earth.

Different passive probes are available, depending on your application. With the included TPP0101/TPP0201 passive probes, the TPS2000B can measure up to 400 Vp-p. However, to meet the safety rating of the TPP0101/TPP0201, the reference lead of the probe must be maintained within 30 VRMS relative to ground. Because of this, the TPP0101/TPP0201 probes are well suited for working on digital and analog circuits in which the maximum voltage never exceeds 30 VRMS.


Input signal and float voltage maximum safety ratings.

Measurements on power conversion electronics usually require probes with higher voltage ratings. Tektronix offers two passive probes with insulation systems specifically designed for making floating measurements. Optional P5122 probes, when coupled with the TPS2000B, are suitable for making measurements on 480 VRMS devices in Category II environments, with a maximum float voltage of up to 600 VRMS relative to earth ground. With the optional P5120 probe the TPS2000B can measure up to 800 Vp-p, with a maximum float voltage within 600 VRMS of ground. With the optional P5120 probe the TPS2000B can measure up to 800 Vp-p with a maximum float voltage within 600 VRMS of ground. The P5120 is the best choice for making AC-coupled ripple measurements on high-voltage DC power supplies.

Please see “Characteristics” for complete safety ratings and specifications.

Selecting the Right Probes for the Job

Scope/Probe (Attenuation)

Maximum Safety Ratings

TPS2000B Viewable Signal

Reference Float Safety Rating*1

Input Signal Safety Rating

On-screen Peak-Peak Voltage (Sinusoid centered at 0 V)

On-screen RMS Voltage (Sinusoid centered at 0 V)

TPS2000B Input (1X)

600 VRMS CAT II

300 VRMS CAT II

40 Vp-p

14.1 VRMS

TPP0101 (100 MHz)

TPP0201 (200 MHz)

30 VRMS

300 VRMS CAT II

400 Vp-p

141 VRMS

P5120 (20X)

600 VRMS CAT II

1000 VRMS CAT II

800 Vp-p

282 VRMS

P5122*2 (100X)

600 VRMS CAT II

1000 VRMS CAT II

2828 Vp-p

1000 VRMS

*1 Passive probe reference leads have no attenuation so any working voltage or over-voltage transients pass straight through to the scope reference. Thus, a passive probe reference float rating can never exceed the scope reference float rating.

*2 The P5122 probe should not be used for AC-coupled measurements on signals with greater than 300 V DC offset. The P5120 is the recommended probe for measuring ripple on high-voltage DC supplies.

Speed the Design and Test of Industrial Power Systems and Circuits

From mobile phones to industrial motor drives, power conversion technology has enabled significant advances in size, performance, and energy efficiency. But even the most basic task of viewing a converter’s input and output is complicated by multiple voltage references. Multiple references also make it challenging to view signals from control circuits and power circuits at the same time. Using ground-referenced oscilloscopes in these applications, without appropriate differential probes, can damage circuits and produce bad measurements. For debugging power conversion electronics, IsolatedChannel technology reduces the risk of damage and unintended circuit interactions.

For performing power system measurements, TPS2PWR1 power application software is available as an option for the TPS2000B. It provides advanced power measurements right on the oscilloscope, at an entry-level price.

For dialing in the performance of switching components, the power application software adds important measurements to the TPS2000B, including automatic switching loss, dv/dt, and di/dt cursor measurements.


Perform three-phase power measurements of variable frequency drives.


Conduct harmonic distortion measurements with TPS2PWR1 software.

For measurements on AC line voltage and for checking the impact on the power distribution system, the power application software shows harmonic content to the 50th harmonic, and provides phase, reactive power consumption, and power factor measurements. With the four-channel TPS2014B or TPS2024B, you can view three-phase voltages or currents.

Two power “bundles” are available, combining probes and measurement software. Each package combines four probes with the TPS2PWR1 power application software, at prices that are lower than if purchased separately. The TPS2PBND combines four P5120 20X passive, high-voltage probes with TPS2PWR1 power application software. The TPS2PBND2 combines four P5122 100X passive, high-voltage probes with the power application software.

Quickly Debug and Characterize Signals with DRT Sampling Technology

Capture elusive glitches – the first time – with Digital Real-Time (DRT) sampling technology.

Characterize a wide range of signal types on up to four channels simultaneously with the TPS2000B Series Digital Real-Time (DRT) sampling technology. This acquisition technology makes it possible to capture high-frequency events, such as glitches and edge anomalies, that eludes other oscilloscopes in its class, so that you can be sure to get an accurate view of your signal.

Easily Analyze and Document Your Measurement Results

Speed documentation and analysis of measurements results with OpenChoice® software and integrated CompactFlash® mass storage.

Quickly reveal signal interference, crosstalk, and the effects of vibration with frequency domain analysis using the TPS2000B Series Fast Fourier Transform (FFT) feature. Then, easily document your measurement results with the integrated CompactFlash® mass storage.

To capture, save, and analyze your measurement results on your PC, the included OpenChoice® PC software can be used. Every TPS2000B Series also ships with a free copy of the special Tektronix Edition of National Instrument’s LabVIEW SignalExpress™ software for basic instrument control, data logging, and analysis. The optional Professional Edition of SignalExpress offers over 200 built-in functions that provide additional signal processing, advanced analysis, sweeping, limit testing, and user-defined step capabilities.

SignalExpress supports the range of Tektronix bench instruments*3, enabling you to connect your entire test bench. You can then access the feature-rich tools packed into each instrument from one intuitive software interface. This allows you to automate complex measurements requiring multiple instruments, log data for an extended period of time, time-correlate data from multiple instruments, and easily capture and analyze your results, all from your PC. Only Tektronix offers a connected test bench of intelligent instruments to simplify and speed debug of your complex design.

*3 To see the full range of Tektronix instruments supported by the Tektronix Edition of NI LabVIEW SignalExpress, visit www.tektronix.com/signalexpress.

Correlate Your Measurements from Bench to Lab to Field*4

Easily correlate your measurements between bench, lab, and field with the highly portable TPS2000B Series.


Enjoy virtually unlimited freedom from an AC power source with hot-swappable batteries.

Use the TPS2000B Series on your bench, in the lab, or in the field, with the industry’s longest continuous battery life – 8 hours and beyond – in a highly mobile package. Enjoy virtually unlimited freedom from an AC power source with hot-swappable batteries.

Optimize Your Productivity

Easily use the oscilloscope even in environments that challenge operation, with features such as analog-style knobs per channel and backlit menu buttons.

The oscilloscope has a front-panel layout that most users will find familiar. Each channel has a dedicated set of scale and position controls. Reduce your measurement time with features like autoset, autorange, automatic measurements, probe check wizard, and context-sensitive help. Backlit menu buttons help you work in a variety of challenging environments – from bright daylight to dimly lit areas.

Performance You Can Count On

In addition to industry-leading service and support, every TPS2000B Series oscilloscope comes backed with a three-year standard warranty.

*4 Please refer to Environmental and Safety specifications.

Characteristics

TPS2000B Series Electrical Characteristics

Feature

TPS2012B

TPS2014B

TPS2024B

Isolated Channels

2

4

4

Bandwidth*5 (MHz)

100

100

200

Sample Rate (GS/s) per Channel

1.0

1.0

2.0

Record Length

2.5k points

Display (1/4 VGA LCD)

Color

Battery Operation

Capacity for two hot-swappable battery packs

One standard battery pack offers 4 hours of battery operation

Optional second battery pack extends battery operation to 8 hours

Continuous battery operation is possible by hot-swapping charged batteries

Automatic Measurements

11

Isolated External Trigger Input (Impedance isolated)

Yes

Vertical Resolution

8 bits (normal or with averaging)

Vertical Sensitivity

2 mV to 5 V/div on all models with calibrated fine adjustment

DC Vertical Accuracy

±3%

Vertical Zoom

Vertically expand or compress a live or stopped waveform

Max Input Voltage (1 MΩ)

300 VRMS CAT II from BNC signal to BNC shell

Float Voltage

600 VRMS CAT II from BNC shell to earth ground

Position Range

2 mV to 200 mV/div ±1.8 V

>200 mV to 5 V/div ±45 V

Bandwidth Limit

20 MHz

Linear Dynamic Range

±5 div

Time Base Range

5 ns to 50 s/div

5 ns to 50 s/div

2.5 ns to 50 s/div

Time Base Accuracy

50 ppm

Input Impedance

1 MΩ ±2% in parallel with 20 pF

Input Coupling

AC, DC, GND

Horizontal Zoom

Horizontally expand or compress a live or stopped waveform

FFT

Standard

RS-232, Centronics-Parallel Ports

Standard

PC Connectivity

Standard

Integrated CompactFlash® Mass Storage

Standard

Power Measurements

Optional package that offers instantaneous power waveform analysis, waveform analysis, harmonics analysis, switching loss, phase angles, dv/dt and di/dt cursors

*5 Bandwidth is 20 MHz at 2 mV/div, all models. For TPS2024B, 200 MHz bandwidth is typical at 5 mV/div. Bandwidth is 200 MHz at 10 mV/div and above, for operating temperatures from 0 °C to 40 °C. Bandwidth is 180 MHz for all V/div settings 10 mV/div and above, for operating temperatures from 0 °C to 50 °C.

Acquisition Modes

Mode

Description

Peak Detect

High-frequency and random glitch capture. Captures glitches as narrow as 12 ns typical using acquisition hardware at all time/div settings from 5 μs/div to 50 s/div

Sample

Sample data only

Average

Waveform averaged, selectable: 4, 16, 64, 128

Single Sequence

Use the Single Sequence button to capture a single triggered acquisition sequence at a time

Scan/Roll Mode

At acquisition time-base settings of ≥100 ms/div

Trigger System (Main Only)

Characteristic

Description

Trigger Modes

Auto, Normal, Single Sequence

Trigger Types

Trigger

Description

Edge (Rising or falling)

Conventional level-driven trigger. Positive or negative slope on any input.

Coupling Selections: AC, DC, Noise Reject, HF Reject, LF Reject

Video

Trigger on all lines or individual line, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM)

Pulse Width (or glitch)

Trigger on a pulse width less than, greater than, equal to, or not equal to a selectable time limit ranging from 33 ns to 10 s

Trigger Source

Characteristic

Description

2-channel Models

CH1, CH2, Ext, Ext/5, Ext/10

4-channel Models

CH1, CH2, CH3, CH4, Ext, Ext/5, Ext/10

Trigger View

Displays trigger signal while trigger view button is depressed.

Trigger Signal Frequency Readout

Provides a frequency readout of the trigger source with 6-digit resolution.

Cursors

Characteristic

Description

Types

Voltage, Time

Measurements

ΔT, 1/ΔT (frequency), ΔV, dv/dt*6, di/dt*6

Measurement System

Characteristic

Description

Automatic Waveform Measurements

Period, Frequency, +Width, –Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, Cycle RMS

Waveform Processing

Characteristic

Description

Operators

Add, Subtract, Multiply, FFT

FFT

Windows: Hanning, Flat Top, Rectangular; 2048 sample points

Sources

   2-channel Models

CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2

   4-channel Models

CH1 – CH2, CH2 – CH1, CH3 – CH4, CH4 – CH3, CH1 + CH2, CH3 + CH4, CH1 × CH2, CH3 × CH4

Autoset Menu

Single-button, automatic setup of all channels for vertical, horizontal, and trigger systems, with undo autoset

Autorange

Allows the user to change test points without resetting the oscilloscope

Autoset Menu for Multiple Signal Types

Signal Type

Autoset Menu Choices

Square Wave

Single Cycle, Multicycle, Rising or Falling Edge

Sine Wave

Single Cycle, Multicycle, FFT Spectrum

Video (NTSC, PAL, SECAM)

Video (NTSC, PAL, SECAM)

Field:

   All, Odd, or Even Line: All or Selectable Line Number

Nonvolatile Storage

Characteristic

Description

Nonvolatile Storage

CompactFlash® up to 2 GB

Reference Waveform Display

Two 2500 point reference waveforms

Waveform Storage

96 or more reference waveforms per 8 MB

Setups

4000 or more front-panel setups per 8 MB

Screen Images

128 or more screen images per 8 MB (the number of images depends on file format selected)

Save All

12 or more Save All operations per 8 MB. A single Save All operation creates 2 to 9 files (setup, image, plus one file for each displayed waveform)

Display Characteristics

Characteristic

Description

Display

1/4 VGA Active TFT Color LCD display

Interpolation

Sin (x)/x

Display Types

Dots, vectors

Persistence

Off, 1 sec, 2 sec, 5 sec, Infinite

Format

YT and XY

I/O Interface

Characteristic

Description

RS-232 Port (Standard)

9-pin DTE

RS-232 Programmability

Full talk/listen modes. Control of all modes, settings, and measurements. Baud rate up to 19,200

Mass Storage CompactFlash® Memory

Accepts any Type 1 CompactFlash® card, up to and including 2 GB (card not included)

Built-in Clock/Calendar

OpenChoice PC Communications Software

Seamless connection from oscilloscope to PC through RS-232.

Transfer and save settings, waveforms, measurements, and screen images.

Includes a Windows desktop data transfer application in addition to convenient Microsoft Word and Excel toolbar add-ins

Printer Port (Standard)

Centronics-type Parallel

   Graphics File Formats

TIFF, PCX (PC Paint Brush), BMP (Microsoft Windows), EPS (Encapsulated Postscript), and RLE

   Printer Formats

Bubble Jet, DPU-411, DPU-412, DPU-3445, Thinkjet, Deskjet, Laser Jet, Epson Dot (9- or 24-pin), Epson C60, Epson C80

   Layout

Landscape and Portrait

*6 Requires TPS2PWR1 power application package.

Environmental and Safety

Characteristic

Description

Temperature

   Operating

0 °C to +50 °C

   Nonoperating

–40 °C to +71 °C

Humidity

TPS2000B Series oscilloscopes are not intended for use in wet or damp conditions

   Operating

High: 50 °C / 60% RH

Low: 30 °C / 90% RH

   Nonoperating

High: 55 °C to 71 °C / 60% RH max wet bulb

Low: 30 °C to 0 °C / <90% RH max wet bulb

Altitude

   Operating

Up to 3,000 m

   Nonoperating

15,000 m

Pollution Degree 2

Do not operate in an environment where conductive pollutants may be present (as defined in IEC61010-1:2001)

Enclosure Rating

   IP30

When the CompactFlash® card and power analysis software are installed (as defined in IEC60529:2001)

Electromagnetic Compatibility

Meets the intent of Directive 89/336/EEC.

Meets or Exceeds: Australian EMC Framework, demonstrated per Emission Standard AS/NZS 2064.1/2

Safety

UL61010-1: 2004. CAN/CSA22.2 No. 1010.1: 2004. EN61010-1: 2001.

Do not float the TPP0101/TPP0201 probe common lead to >30 VRMS. Use the P5122, P5120 (floatable to 600 VRMS CAT II) or similarly rated passive, high-voltage probe, or an appropriately rated high-voltage, differential probe when floating the common lead above 30 VRMS

CAT Ratings

Overvoltage Categories

Category

Examples of Products in this Category

CAT III

Distribution-level mains, fixed installation

CAT II

Local-level mains, appliances, portable equipment

CAT I

Signal levels in special equipment or parts of equipment, telecommunications, electronics

 

Materials - TPSBAT battery contains less than 8 grams equivalent Lithium.

Physical Characteristics

INSTRUMENT

Dimensions

mm

in.

Width

336.0

13.24

Height

161.0

6.33

Depth

130.0

5.10

Weight

kg

lb.

Instrument Only

2.7

6.0

   with 1 battery

3.2

7.0

   with 2 batteries

3.7

8.0

INSTRUMENT SHIPPING

Package Dimensions

mm

in.

Width

476.2

18.75

Height

266.7

10.50

Depth

228.6

9.00

Ordering Information

TPS2012B, TPS2014B, TPS2024B

Digital Storage Oscilloscopes.

Standard Accessories

Accessory

Description

Probes

TPP0101 100 MHz, 10X passive probe for TPS2012B and TPS2014B; TPP0201 200 MHz, 10X passive probe for TPS2024B; one probe per channel standard

Battery (1)

Lithium-ion battery with fuel gauge for 4-hour battery life. Two batteries required for 8 hours of continuous battery operation

USB to RS-232 Cable

A cable to connect to a PC USB port through the RS-232 port on the backside of the scope

OpenChoice® PC Connectivity Software

A collection of programs that enable fast and easy communication between MS Windows PCs and TPS2000B Series oscilloscopes

NI SignalExpress™ Tek Edition Software

A program that enables easy communication between PC-based NI SignalExpress software and TPS2000B Series oscilloscopes

Documentation

User Manual (please see below for available language options)

AC Adapter with Power Cord

 

NIM/NIST-Traceable Certificate of Calibration

 

Front Protective Cover

 

Recommended Accessories

Accessory

Description

TPS2PBND2

Power bundle for TPS2000B oscilloscopes. Includes (4) P5122 passive, 100X high-voltage probes and TPS2PWR1 power measurement and analysis software

TPS2PWR1

Power measurements application package. Instantaneous power waveform analysis, waveform analysis, harmonics analysis, switching loss, phase angles, dv/dt and di/dt cursors

WSTRO

WaveStar software; Microsoft Windows application for waveform capture, analysis, documentation, and control from your PC. Provides enhanced oscilloscope data measurement, analysis, remote setup and charting features

TPSBAT

Additional battery

TPSCHG

Battery charger

AC2100

Soft case for carrying instrument

HCTEK4321

Hard case for carrying instrument (requires AC2100)

077-0447-xx

Service Manual – English only

077-0444-xx

Programmer Manual – English only

Recommended Probes

Accessory

Description

A621

2000 A, 5-50 kHz AC current probe/BNC

A622

100 A, 100 kHz AC/DC current probe/BNC

P5122*2

200 MHz passive 100X high-voltage probe

P5205

High-voltage active differential probe (1300 Vp-p, 100 MHz) (1103 power supply required)

P5210

High-voltage active differential probe (5600 Vp-p, 50 MHz) (1103 power supply required)

CT2

2.5 A, 200 MHz AC current probe

TCP202

15 A, 50 MHz AC/DC current probe (1103 power supply required)

TCP303/TCPA300

150 A, 15 MHz AC/DC current probe/amplifier

TCP305/TCPA300

50 A, 50 MHz AC/DC current probe/amplifier

TCP312/TCPA300

30 A, 100 MHz, DC/AC current probe/amplifier

TCP404XL/TCPA400

500 A, 2 MHz AC/DC current probe/amplifier

*2 The P5122 probe should not be used for AC-coupled measurements on signals with greater than 300 V DC offset. The P5120 is the recommended probe for measuring ripple on high-voltage DC supplies.

International Power Plugs

Option

Description

Opt. A0

North America power

Opt. A1

Universal Euro power

Opt. A2

United Kingdom power

Opt. A3

Australia power

Opt. A5

Switzerland power

Opt. A6

Japan power

Opt. A10

China power

Opt. A11

India power

Opt. A12

Brazil power

Opt. A99

No power cord or AC adapter

Accessory Cables

Cable

Description

012-1241-xx

RS-232, 9-Pin Female to 25-Pin Male, 4.6 m (15 ft.), for modems

012-1651-xx

RS-232, 9-Pin Female to 9-Pin Female, null modem, for computers

012-1380-xx

RS-232, 9-Pin Female to 25-Pin Female, null modem, for computers

012-1214-xx

Centronics, 25-Pin Male to 36-Pin Centronics, 2.4 m (8 ft.), for parallel printer interfaces

International User Manual Language Options

Option

Description

Translated front-panel overlay included with its respective user manual.

Opt. L0

English (071-1441-xx)

Opt. L1

French (071-1442-xx)

Opt. L2

Italian (071-1443-xx)

Opt. L3

German (071-1444-xx)

Opt. L4

Spanish (071-1445-xx)

Opt. L5

Japanese (071-1446-xx)

Opt. L6

Portuguese (071-1447-xx)

Opt. L7

Simplified Chinese (071-1448-xx)

Opt. L8

Traditional Chinese (071-1449-xx)

Opt. L9

Korean (071-1450-xx)

Opt. L10

Russian (071-1451-xx)

Service Options

Option

Description

Opt. SILV200

Standard Warranty Extended to 5 Years

 

Warranty Information

Three-year warranty covering all labor and parts, excluding probes and accessories.

 

Speed Product Development with Best-in-Class Price/Performance

The extensive Tektronix portfolio of proven, state-of-the-art stimulus, probing, acquisition, and analysis tools simplify and speed each phase of product design – from power-on and verification, through debug and validation, to characterization and test – to enable you to race products to your customers when they need them, if not before.

 

Tektronix Support Completes the Solution

Anytime you need support, anywhere in the world, depend on Tektronix Support to give you the lowest possible exposure to inconvenience, delay, or disruption of operations. www.tektronix.com/support

  • Unsurpassed technical expertise and experience with 24-hour response to technical questions
  • Industry-leading turnaround service time
  • 90-day unconditional service warranty
  • No fine print, no exclusions, no surprises
  • Global support in more than 50 countries
Digital Storage Oscilloscopes
STARTING PRICES:

TPS2000 : $2,990

Configure & Quote

Select Model

x
MORE INFORMATION:

Select Service

X

STAY INFORMED: Events & Seminars | Email Newsletter | Follow Tektronix on Twitter Follow Tektronix on Facebook Follow Tektronix on Google+ Follow Tektronix on Youtube

MEMBERSHIP: Create an account for exclusive membership privileges. Learn More | Login