Current Probes

A621 • A622

Features & Benefits

A621
  • AC – 5 Hz to 50 kHz
  • 100 mA to 2000 A Peak
  • For DMMs and Oscilloscopes
  • Clamp On
A622
  • AC/DC – 100 kHz
  • 50 mA to 100 A Peak
  • For DMMs and Oscilloscopes

Applications

  • Motor Drives
  • Inverters
  • Power Supplies
  • Avionics

A622

 

 

 

 

 

The A600 Series current probes high-value current probes are specifically designed to support measurements you make with your TekMeter® or oscilloscope*1.

A621 2000 Amp AC Current Probe/BNC

This industrial-style clamp-on probe has a BNC connector and may be used with a shrouded banana plug adapter*1 so it can be used on digital multimeters, TekMeter, and scopes. The A621 can measure AC currents from 100 mA to 2000 A peak over a frequency range of 5 Hz to 50 kHz. It provides a 1 mV, 10 mV, or 100 mV output for each Amp measured.

A622 100 Amp AC/DC Current Probe/BNC

This "long nose" style clamp-on probe uses a Hall Effect current sensor to provide a voltage output to oscilloscopes. It has a BNC connector and may be used with a shrouded banana plug adapter*1 so it can also be used on digital multimeters, TekMeter, and scopes. The A622 can measure AC/DC currents from 50 mA to 100 A peak over a frequency range of DC to 100 kHz. It provides 10 mV or 100 mV output for each Amp measured.

Recommended Products

TPS2000, TDS1000B, TDS2000C, and TDS3000C Series oscilloscopes and DMM4020*1, DMM4040*1, and DMM4050*1 digital multimeters.

*1 For instruments with banana jack inputs, Tektronix part number 012-1450-00 Female BNC to banana lead adapter is required.

 

Characteristics

Characteristic

A621

A622

Frequency Range

5 Hz to 50 kHz

DC to 100 kHz

Max Input Current

2000 A peak

100 A peak

Output

1 mV/A, 10 mV/A, 100 mV/A

10 mV/A, 100 mV/A

Max Conductor Diameter

54 mm (2.13 in.)

11.8 mm (0.46 in.)

Termination

BNC*1

BNC*1

Max Bare-wire Voltage

600 V (CAT III)

600 V (CAT III)

Safety

UL3111-2-032

CSA1010.2.032

EN61010-2-032

IEC61010-2-032

UL3111-2-032

CSA1010.2.032

EN61010-2-032

IEC61010-2-032

*1 For instruments with banana jack inputs, Tektronix part number 012-1450-00 Female BNC to banana lead adapter is required.

 

Ordering Information

A621

2000 A AC Current Probe/BNC.

A622

100 A AC/DC Current Probe/BNC.

Service

Option

Description

Opt. R3

Repair Service 3 Years.

Opt. R5

Repair Service 5 Years.

Recommended Accessories

Accessory

Description

012-1450-xx

Adapter, lead; discrete – MLD, 2, 18 AWG, dual insul, BNC, female X 4 mm dual insul; banana jack X dual insul plug, shield banana

Select Service

X

STAY INFORMED: Events & Seminars | Email Newsletter | Follow Tektronix on Twitter Follow Tektronix on Facebook Follow Tektronix on Google+ Follow Tektronix on Youtube