
Figure 1b. - Optional TDS3BAT Battery Pack.
However, AC power is not always available where you want to make
oscilloscope measurements. The TDS3000 Series optional battery pack (TDS3BAT)
allows you to operate the oscilloscope without the need for AC power. Observing
the following precautions will provide safe operation of the TDS3000 Series
oscilloscopes when battery powered.
Figure 2 shows the warning that is displayed on the TDS3000 Series screen
when in battery operation mode.
Figure 2. - Warning displayed by TDS3000 Series when operating in battery mode.
This warning applies to circuits that have voltages of greater than 30 VRMS
(>42 Vpk). If the circuit under test has voltages that exceed 30
VRMS (>42 Vpk), the TDS3000 Series chassis needs to
be connected to earth ground using the grounding wire provided with the
instrument (Battery Pack) to prevent electrical shock to the operator.
If you do not attach the grounding wire, you are not protected against
electrical shock if you connect the oscilloscope to a hazardous voltage.
WARNING
Such electrical shock could result in serious personal
injury or loss of life.
While in battery operation and following environmental specification limits
for the TDS3000 Series, it is safe to "float" the "signal common"
for making measurements provided you do not connect a signal greater than 30 VRMS
(>42 Vpk) from earth ground to either the probe tip or common
lead. For measurements where higher voltages {>30 VRMS, >42 Vpk)
are present, the instrument's chassis must be connected to earth ground using
the supplied grounding wire to prevent electrical shock to the operator.
WARNING
If there is any doubt whether more than 30 VRMS
is present or not, the supplied grounding wire should always be used and
floating measurements SHOULD NOT BE ATTEMPTED! Be aware that hazardous voltages
may exist in unexpected places due to faulty circuitry in the device-under-test.
It's important to remember that the "signal commons" for all
channels are at the same potential and are NOT independent. Ensure that all
probe common leads are connected to the same voltage or common point.
Do not connect a grounded device, such as a printer or computer, to the
oscilloscope unless the oscilloscope's grounding wire is connected to earth
ground!
In order to perform differential measurements with independent "signal
grounds," true differential probes such as the P5205 or P5210 must be used.
Safety Engineering Principles
Monolithic Isolation Amplifiers
|