Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

Real-Time Spectrum Analyzers

RSA3303A • RSA3308 • WCA230A • WCA280A

The products on this datasheet are no longer being sold by Tektronix.
View Tektronix Encore for reconditioned test equipment.
Check support and warranty status for these products.

RSA3303A • RSA3308 • WCA230A • WCA280A

Features & Benefits

  • Trigger
    • Tektronix' Exclusive Frequency Mask Trigger Makes Event-based Capture of Transient RF Signals Easy By Triggering on Any Change in the Frequency Domain
  • Capture
    • All Input Signals up to 15 MHz*1 Spans Are Seamlessly Captured Into Memory
    • Long Record Length Enables Complete Analysis Over Time Without Making Multiple Acquisitions
    • Interfaces With TekConnect® Probes for RF and Baseband Probing
  • Analyze
    • Gain a Unique Understanding of Time-varying Transient RF Signals
    • Spectrogram Provides a Revealing Picture of RF Signal Frequency and Amplitude Behavior Over Time – Not possible With a Swept Spectrum Analyzer
    • Multi-domain Analysis Enables Fast, Complete Signal Analysis in Frequency, Time, Code and Modulation Domains Without Making Multiple Measurements
    • Simple Capture and Analysis on RFID Interrogator and Response Signals
    • Comprehensive Pulse Measurement Suite
    • General Purpose Digital Modulation Analysis
    • Spectrum Analyzer View for Traditional Wide Band Signal Analysis
    • 3G Measurement Versatility with W-CDMA, cdma2000, 1X EVDO, HSUPA, HSDPA, TD-SCDMA RF and Modulation Analysis (WCA200A Only)
    • Signal Source Analysis Simplifies Phase Noise, Jitter and Frequency Settling Measurements

Applications

  • System Integration of 3G, and Other RF Systems
  • Radar and Pulsed RF Signal Characterization
  • RFID System Development and Troubleshooting
  • General Purpose Phase Noise and Jitter Signal Analysis
  • Characterization of Interfering or Unknown Signals in Spectrum Monitoring and Surveillance
  • Troubleshooting RF Components, Modules, or Systems
  • Getting Answers to Elusive EMI Diagnostic Problems

*1 20 MHz bandwidth at Baseband.

Trigger, Capture, Analyze Radar, 3G, or Other Time-Varying RF Signals

A000_5647-N

High-resolution spectrogram reveals transient signal behavior that translates to rapid problem solving. Here, 500 kHz sidebands are revealed as part of the transient behavior of a hopping signal as it switches frequencies.

   

A000_5648-N

Time-correlated, multi-domain view provides a new level of insight into design or operational problems not possible with conventional analysis solutions.

 

Get fast resolution to complex problems with Trigger, Capture and Analysis tools

See the frequency and amplitude of your RF signal change over time in a single view. With only a single acquisition, the RSA3300A and WCA200A Series Real-Time Spectrum Analyzers (RTSA) capture a continuous time record of changing RF events, and enables time-correlated analysis in the frequency, time, and modulation domains. You get the functionality of a vector signal analyzer, a wide band spectrum analyzer, plus the unique trigger-capture-analyze capability of RTSA — in one, transportable package.

Trigger

Patented 15 MHz bandwidth Frequency Mask Trigger (FMT) makes it easy to capture transient, low duty-cycle or other difficult to capture signals. An FMT mask is simply configured using a mouse and it can be set up for one or many frequency bands within an analysis span. FMT can monitor for signal appearance/disappearance, or change in amplitude, frequency, bandwidth, spectral shape, and more—all while the instrument user is working on another task. A Power Trigger, working in the time domain and at any Real-Time analysis span, can be armed to monitor for a user-set power threshold to be crossed during a moment in time. A power detector determines total power of all signals in a span which is compared to the user-set threshold.

Capture

Capture once—make multiple measurements as needed. All signals in a Real-Time analysis span—including transients, low duty-cycle and other difficult-to-measure events—are captured together into deep memory where signal data can be accessed at the user's convenience. Record lengths vary depending on span selected—up to 2.56 s at 15 MHz span, 40.96 s at 1 MHz span, or 4096 s at 10 KHz span with Deep Memory Opt. 02. Real-Time capture of small signals is enhanced by –74 dBc third order IM, plus very good phase noise performance and sensitivity. A solid performance front-end serves not only Real-Time and wide band Spectrum Analysis modes, but also on-board vector signal analysis functionality.

The RTPA2A Real Time Probe Adapter extends the capabilities of the Real-Time Spectrum Analyzers (RTSA) by offering additional tools to make debugging today’s high-performance electrical designs easier. Using the RTPA2A with Tektronix RTSA, design engineers can benefit from Tektronix' industry-leading active and differential probes to measure signals on SMD pins, or other challenging circuit features.

Analyze

Time-correlated multi-domain analysis provides engineers with unique insight into time-varying signal behavior, resulting in fast analysis and problem solving. Time-correlated measurements can be made across the frequency, time, and modulation domains. The analysis display called Spectrogram has the ability to overlap individual spectra as close as 40 ns, providing an intuitive view of signal changes over time, ideal for such things as frequency hopping, pulsed signals, modulation switching, settling time, bandwidth changes, relative timing of appearing, and intermittent signals. The RTSAs introduce analysis capabilities that advance productivity for engineers working on components or in RF system design, integration and performance verification, or operations engineers working in networks, spectrum monitoring, or surveillance.

Example Applications Benefiting from Key RSA3300A and WCA200A Capabilities

Analysis Feature

RF Communications Systems

Cellular Devices

Radar, Pulsed Signal Transmission

Surveillance, Spectrum Monitoring

RFID

Hi-res Spectrogram

X

X

X

X

X

Multi-domain Correlation

X

X

X

X

X

Cellular Standards Analysis (Multiple Options)

 

X

 

X

 

Advanced Measurement Suite (Opt. 21)

X

 

X

X

X

AM, FM, PM Analysis

X

   

X

X

 

Pulsed RF Signal Analysis

X

 

X

X

 

Pulse Spectrum

X

X

X

X

 

AM/AM, AM/PM and 1 dB compression (Opt. 21)

X

X

X

   

   

TekConnect® Probe Adapter for Real-Time Spectrum Analyzers

The RTPA2A Real Time Probe Adapter extends the capabilities of the Real-Time Spectrum Analyzers (RTSA) by offering additional tools to make debugging today’s high-performance electrical designs easier. Using the RTPA2A with Tektronix RTSA, design engineers can benefit from Tektronix' industry-leading active and differential probes to measure signals on SMD pins or other challenging circuit features.

 

Characteristics

Trigger-related

Trigger Mode - Free run (triggered by acquisition), Triggered (Triggered by event), Single, or Continuous

Trigger Event Source - Power (span BW), Frequency Mask (Opt. 02), External

Pre-/Post-Trigger Setting - Trigger position settable within 0 to 100% of total acquisition length

Trigger Marker Position Timing Uncertainty (Power and External Trigger) -  ±2 sample points

FREQUENCY MASK TRIGGER (Opt. 02)

Mask Resolution - 1 bin

Level Range - 0 dBfs*2 to –60 dBfs at 10 dB/div vertical scale

Bandwidth -

Up to 15 MHz: Start frequency ≥20 MHz

Up to 20 MHz: Start frequency <20 MHz

 

Mask Shape - User-defined

Minimum Horizontal Mask Setting Resolution - <0.2% of span

Uncertainty - ±2 frames

POWER TRIGGER

Level Range - 0 dBfs to –40 dBfs

EXTERNAL TRIGGER

Threshold Voltage - –1.5 V to +1.5 V

Threshold Voltage Setting Resolution - 0.1 V

Input Impedance - >2 kΩ

TRIGGER OUTPUT

Voltage (Output Current <1 mA) - HIGH: >2.0 V; LOW: <0.4 V

*2 dBfs: dB relative to full scale

Capture-related

Real-Time Capture Bandwidth - 15 MHz RF; 20 MHz baseband; 20 MHz using Opt. 03 IQ inputs

A/D Converter - 51.2 MS/s, 14 bits

Minimum Acquisition Length in RTSA/Time/Demod Modes - 1024 samples

Maximum Acquisition Length in RTSA/Time/Demod Modes - 16,384,000 samples; 65,636,000 samples, Opt. 02

Acquisition Length Setting Resolution in RTSA/Time/Demod Modes -

Acquisition Memory Size - 16.4 MSamples; 65.6 M Samples, Opt. 02

Block Size (number of frames) - 1 to 16,000; 1 to 64,000, Opt. 02

Memory Depth (time) and Maximum Time Resolution

Span

Sample Rate (For I and Q)

Record Length

Record Length (Opt. 02)

Spectrum Frame (Time)

Max Time (Resolution)

20 MHz (Baseband)

25.6 MS/s

0.64 s

2.56 s

40 μs

40 ns

15 MHz

25.6 MS/s

0.64 s

2.56 s

40 μs

40 ns

10 MHz

12.8 MS/s

1.28 s

5.12 s

80 μs

80 ns

5 MHz

6.4 MS/s

2.56 s

10.24 s

160 μs

160 ns

2 MHz

3.2 MS/s

6.4 s

20.48 s

320 μs

320 ns

1 MHz

1.6 MS/s

12.8 s

40.96 s

640 μs

640 ns

500 KHz

800 kS/s

25.6 s

81.92 s

1.28 ms

1.25 μs

200 KHz

320 kS/s

64 s

204.8 s

3.2 ms

3.2 μs

100 KHz

160 kS/s

128 s

409.6 s

6.4 ms

6.4 μs

50 KHz

80 kS/s

256 s

819.2 s

12.8 ms

12.8 μs

20 KHz

32 kS/s

640 s

2048 s

32 ms

32 μs

10 KHz

16 kS/s

1280 s

4096 s

64 ms

64 μs

5 KHz

8 kS/s

2560 s

8192 s

128 ms

128 μs

2 KHz

3.2 kS/s

6400 s

20480 s

320 ms

320 μs

1 KHz

1.6 kS/s

12800 s

40960 s

640 ms

640 μs

500 Hz

800 kS/s

25600 s

81920 s

1.28 s

1.28 ms

200 Hz

320 S/s

64000 s

204800 s

3.2 s

3.2 ms

100 Hz

160 S/s

128000 s

409600 s

6.4 s

6.4 ms

Analysis-related

Measurement Functions by Mode

Mode

Measurements

SA

Channel Power, Adjacent Channel Power Ratio, Occupied Bandwidth, Emission Bandwidth, Carrier-to-Noise Ratio, Carrier Frequency

RTSA

Channel Power, Adjacent Channel Power Ratio, Occupied Bandwidth, Emission Bandwidth, Carrier-to-Noise Ratio, Carrier Frequency

Time

IQ vs. Time, Power vs. Time, Frequency vs. Time, CCDF, Crest Factor Pulse Measurements: Pulse Width, Pulse Peak Power, On/Off Ratio, Pulse Ripple, Pulse Repetition Interval, Duty Cycle, Pulse-Pulse Phase, Channel Power, OBW, EBW, Frequency Deviation (Min pulse length, 20 samples; Max pulse length, 360,000 samples)

Analog Demod

IQ vs. Time, AM Depth, FM Deviation, PM Deviation, Pulse Spectrum

Views by Mode

Mode

Views

SA

Spectrum

SA/Spectrogram

Spectrum, Spectrogram

RTSA

Spectrum, Spectrogram

Time

Overview: Power vs. Time, Spectrogram

Sub-view: Spectrum

Main view: Measurement Result

Analog Demod

Overview: Power vs. Time, Spectrogram

Sub-view: Spectrum

Main view: Measurement Result

A000_3248-N

Screen layout, to identify analysis view locations as described in the above chart

 

MEASUREMENT SPEED

Screen Update Rate - 2 MHz Span, Auto RBW: 19.4/s

Remote Measurement Rate and GPIB Transfer Rate (2 MHz span, auto RBW, spectrum data) - 1.87 waveforms/s, or 6,000 samples/s

RF Center Frequency Switching Time -

<10 ms for 10 MHz frequency change

500 ms for 3 GHz frequency change

 

TRACES, DISPLAYS, DETECTORS

Traces - Two traces, Spectrum Analyzer mode

Displays - Up to three time-correlated, user-selected displays

Detector - RMS

Trace Types - Normal (RMS), Average, Max Hold, Min Hold

Display Detection - Max, Min, Max/Min

MODULATION ANALYSIS

Analog

AM

Minimum Input Level - –40 dBfs, typical

PM

 

Minimum Input Level - –40 dBfs, typical

PM Scale, Max, Min - ±180°

FM

 

Minimum Input Level - –40 dBfs, typical

Range - ±Span/2 from center frequency

DEMODULATION ACCURACY

Analog

AM (–10 dBfs signal, input at CF, 10 to 60% modulation depth) - ±2%

PM (–10 dBfs signal, input at CF) - ±3°

FM (–10 dBfs signal, input at CF) - ±1% of span

RF Performance

FREQUENCY

Frequency Range -

RSA3308A/WCA280A: DC to 8 GHz

RSA3303A/WCA230A: DC to 3 GHz

 

Center Frequency Setting Resolution - 0.1 Hz

Frequency Marker Readout Accuracy, Baseband - ±(RE*3 × MF*4 + 0.001 × Span + 0.2) Hz

Frequency Marker Readout Accuracy, RF - ±(RE × MF + 0.001 × Span + 2) Hz

Span Accuracy - ±1 bin

RBW Filter Bandwidth Accuracy - 0.1%

Reference Frequency -

Aging per day - 1 × 10-9 (after 30 days of operation).

Aging per year - 1 × 10-7 (after 30 days of operation).

Temperature Drift - 1 × 10-7 (10 to 40 °C).

Total Frequency Error - 2 × 10-7 (within one year of calibration).

Reference Output Level -  >0 dBm.

External Reference Input - 10 MHz, –10 dBm to + 6 dBm.

Frequency Span -

Range, Spectrum Analyzer Mode -

50 Hz to 3 GHz (Start Frequency ≥40 MHz)

0 Hz to 40 MHz (Stop Frequency <40 MHz)

 

Range, Real-Time Spectrum Analyzer Mode -

100 Hz to 15 MHz (RF)

0 Hz to 20 MHz (Baseband)

 

Resolution Bandwidth Range - 1Hz to 10MHz, automatically selected or user-defined

Accuracy - Within 6.0% ±0.1%

Shape Characteristic - Gaussian with <5:1 shape factor (3:60 dB)

Rectangular, Nyquist, Root Nyquist shapes may also be selected.

 

*3 RE: Reference Frequency Error

*4 MF: Marker Frequency (Hz)

Minimum Settable RBW (Extended Resolution ON)

Frequency

RBW

Span >2 GHz

100 kHz

1 GHz <Span ≤2 GHz

50 kHz

500 MHz <Span ≤1 GHz

20 kHz

20 MHz <Span ≤500 MHz

10 kHz

500 kHz <Span ≤20 MHz

1 kHz

200 kHz <Span ≤500 kHz

500 Hz

100 kHz <Span ≤200 kHz

200 Hz

50 kHz <Span ≤100 kHz

100 Hz

20 kHz <Span ≤50 kHz

50 Hz

10 kHz <Span ≤20 kHz

20 Hz

5 kHz <Span ≤10 kHz

10 Hz

2 kHz <Span ≤5 kHz

5 Hz

1 kHz <Span ≤2 kHz

2 Hz

100 Hz <Span ≤1 kHz

1 Hz

Noise Bandwidth Range, RTSA Mode - 313.18 mHz to 400.87 kHz

FFT Performance -

Number of samples per frame - 64 to 8192 (65,536 samples per frame, extended resolution).

Window types - Rectangular, Parzen, Welch, Sine-Lobe, Hanning, Sine-cubed, Sine-to-the-4th, Hamming, Blackman, Rosenfield, Blackman-Harris 3A, Blackman-Harris 3B, Blackman-Harris 4A, Blackman-Harris 4B, FlatTop.

STABILITY

Noise Sidebands, dBc/Hz

Offset

At 1 GHz CF

At 2 GHz CF

At 6 GHz CF

   

Spec

Typical

Spec

Typical

Spec

Typical

1 kHz

-100

-103

-96

-99

-87

-90

10 kHz

-105

-108

-104

-107

-104

-107

20 kHz

-105

-108

-105

-108

-105

-108

30 kHz

-105

-108

-105

-108

-105

-108

100 kHz

-112

-115

-112

-115

-112

-115

1 MHz

-132

-135

-132

-135

-128

-131

5 MHz

-135

-138

-135

-138

-130

-133

7 MHz

-135

-138

-135

-138

-130

-133

Residual FM - 2 Hz, pk-pk, typical.

AMPLITUDE

Measurement Range - Displayed average noise level to MAX safe input.

Input Attenuator Range -

RF/Baseband Input - 0 dB to 50 dB, 5 dB step.

IQ Input (Opt. 03) - 0 dB to 30 dB, 10 dB step.

Input Attenuator Setting Uncertainty - ±0.5 dB (at 50 MHz).

Maximum Safe Input Level -

Average Continuous (RF band, RF ATT ≥10 dB) - +30 dB.

MAX DC Voltage -

±0.2 V, RF

±5 V, Baseband

±5 V, IQ input w/ Opt. 03

 

Log Display Range - 10 μdB/div to 10 dB/div

Linear Display Scale - 10 divisions

Linear Display Units - dBm, dBμV, V, Watts, Hz for FM Demod, Degrees for PM Demod

Marker Readout Resolution, Log - 0.01 dB

Marker Readout Resolution, Linear - 0.001 μV

Frequency Response, 20 °C to 30 °C, RF ATT ≥10 dB

Frequency

Spec

Typical

100 kHz — 40 MHz

±0.5 dB

±0.3 dB

40 MHz — 3.0 GHz

±1.2 dB

±0.5 dB

3.0 GHz — 6.5 GHz*5

±1.7 dB

±1.0 dB

5 GHz — 8 GHz*5

±1.7 dB

± 1.0 dB

Absolute Amplitude Accuracy at Calibration Point (Baseband at 25 MHz, -10 dBm signal, 0 dB ATT, 20 °C to 30 °C) - ±0.3 dB

Absolute Amplitude Accuracy at Calibration Point (RF at 100 MHz, -20 dBm signal, 0 dB ATT, 20 °C to 30 °C) - ±0.5 dB

Reference Level Setting Range -

1 dB step, RF, -50 dBm to +30 dBm

5 dB step, Baseband, -30 dBm to +20 dBm

5 dB step, IQ, -10 dBm to +20 dBm

 

Reference Level Accuracy (-10 dBm to -50 dBm at 100 MHz, 10 dB ATT, 20 °C to 30 °C) - ±0.2 dB

Level Linearity in Display Range - ±0.2 dB, spec; ±0.12 dB, typical

SPURIOUS RESPONSE

1 dB Compression (RF ATT = 0 dB, 2 GHz CF) - +2 dBm

3rd Order Inter-modulation Distortion (Ref Level = +5 dBm, RF ATT: adjusted for optimum, total signal power = -7 dBm, CF = 2 GHz) - –74 dBc

2nd Harmonic Distortion (-30 dBm tone at input mixer, 10 MHz to 1750 MHz) - –56 dBc, typical.

Displayed Average Noise Level, Specified, dBm/Hz

Frequency

Spec

10 MHz

–151

2 GHz

–150

3 GHz

–150

7 GHz*5

–142

Displayed Average Noise Level, Typical, dBm/Hz

Frequency

Typical

1 kHz –10 kHz

–144

10 kHz –10 MHz

–151

10 MHz –100 MHz

–151

100 MHz –1 GHz

–150

1 GHz – 2 GHz

--150

2 GHz –3 GHz

–150

3 GHz – 5 GHz*5

–142

5 GHz – 8 GHz*5

–142

Residual Response

Frequency

Spec

1 to 40 MHz (Span =20 MHz, Ref Lvl = –30 dBm, RBW –100 kHz)

–93 dBm

0.5 to 3.5 GHz (Span =3 GHz, Ref Lvl = –30 dBm, RBW =100 kHz)*5

–90 dBm

3.5 to 6.5 GHz (Span =3 GHz, Ref Lvl = –30 dBm, RBW =100 kHz)*5

–85 dBm

3.5 to 8 GHz (Span =3 GHz, Ref Lvl = –30 dBm, RBW =100 kHz)*5

–85 dBm

Spurious Response with Signal

Frequency

Spec

0 MHz (Span =10 MHz, Ref Lvl =0 dBm, RBW –50 kHz, Signal Frequency =25 MHz, Signal Level = –5 dBm)

–73 dBc

2 GHz (Span =10 MHz, Ref Lvl =0 dBm, RBW –50 kHz, Signal Frequency =2 GHz, Signal Level = –5 dBm)

–73 dBc

5 GHz (Span =10 MHz, Ref Lvl =0 dBm, RBW –50 kHz, Signal Frequency =5 GHz, Signal Level = –5 dBm)*5

–70 dBc

7 GHz (Span =10 MHz, Ref Lvl =0 dBm, RBW –50 kHz, Signal Frequency =7 GHz, Signal Level = –5 dBm)*5

–70 dBc

*5 Frequency >3 MHz available on RSA3308A, WCA280A only

INPUTS AND OUTPUTS

Front Panel

Input Connectors - N type, RF/Baseband; BNC type, IQ, Opt. 03

Input Impedance - 50 Ω

VSWR, RF ATT >10 dB

Frequency

Spec

Typical

300 kHz –10 MHz

<1.4:1

10 MHz – 3 GHz

<1.3:1

2.5 GHz

<1.4:1

7.5 GHz

<1.8:1

Preamp Power Connector - LEMO 6 pin connector - Pin 1: NC; Pin 2: ID1; Pin 3: ID2; Pin 4: -12 V; Pin 5: GND; Pin 6: +12 V.

External Preamp (Opt. 1A) - 100MHz - 3GHz, 20 dB gain, 6.5 dB Noise Figure at 2 GHz (Typical)

Rear Panel

10 MHz REF OUT - 50 Ω, BNC, >–3 dBm

10 MHz REF IN - 50 Ω, BNC, –10 dBm to +6 dBm

EXT TRIG IN - Ext Trig, BNC, High: 1.6 to 5.0 V, Low: 0 to 0.5 V

GPIB Interface - IEEE 488.2

TRIGGER OUT - 50 Ω, BNC, High: >2.0 V, Low: <0.4 V (output current 1 mA)

Side Panel

LAN Interface - Ethernet 10/100Base-T (Standard)

Serial Interface - USB 1.1, 2 ports

VGA Output - VGA compatible, 15 DSUB

General Characteristics

Temperature -

Operating: +10 °C to +40 °C.

Storage: –20 °C to +60 °C.

 

Warm-up Time - 20 min.

Altitude -

Operating: up to 3000 m (10,000 ft.).

Non-operating: up to 12,000 m (40,000 ft.).

 

Safety and EMI Compatibility -

UL 61010-1; CSA C22.2 No. 61010-1-04; IEC61010, Second Edition (Self Declaration).

Low Voltage Directive 73/23/EEC, amended by 93/68/EEC; EN61010-1: 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use.

EC Council EMC Directive 89/336/EEC, amended by 93/68/EEC.

EN61326-1: 1997 Product Family Standard for Electrical Equipment for Measurement, Control, and Laboratory Use-EMC Requirements.

Electromagnetic Compatibility Framework:1992 AS/NZS 2064.1/2(Industrial, Scientific, and Medical Equipment).

 

Power Requirements - 100 VAC to 240 VAC, 47 Hz to 63 Hz.

Power Consumption - 400 VA max.

Data Storage - Internal HDD (40 GB), USB port, FDD.

Weight, without options - 19 kg, 42 lbs.

Dimensions -

Without bumpers and feet: 215 mm (H) × 425 mm (D) × 425 mm (W).

With bumpers and feet: 238 mm (H) × 470 mm (D) × 445 mm (W).

 

Calibration Interval - 1 year.

Warranty - 1 year.

GPIB - SCPI-compatible.

 

Ordering Information

WCA230A, WCA280A

Real-Time Spectrum Analyzer WCA230A, DC – 3 GHz

Real-Time Spectrum Analyzer WCA280A, DC – 8 GHz.

Includes: User manual, Programmer's manual, Power cord, BNC-N adapter, USB Keyboard and Mouse.

Product Options*6

Opt. 02 - 65.5 MSample Deep Memory, Frequency Mask Trigger.

Opt. 03 - IQ, Differential IQ Inputs.

Opt. 23 - W-CDMA Uplink Analysis.

Opt. 24 - GSM/EDGE Analysis.

Opt. 25 - CDMA 1X Forward/Reverse Link Analysis.

Opt. 26 - 1X EVDO Forward/Reverse Link Analysis.

Opt. 27 - 3GPP Release 5 Downlink (HSDPA) Analysis.

Opt. 28 - TD-SCDMA Analysis.

Opt. 40 - GPP Release 6 (HSUPA) Analysis*7.

Opt. 1A - External Preamp, 100 MHz – 3 GHz, 20 dB gain, 6.5 dB Noise Figure at 2 GHz (typical).

Opt. 1R - Rack mount kit.

Opt. SASW - USB Stand Alone Software Key.

*6 Specifications for Options 21 through 40 can be found in the Real-Time Spectrum Analyzer Software Options datasheet, on www.tektronix.com/RSA.

*7 Option 23 and Option 27 are required for 3GPP Release 6 (HSUPA) Analysis in addition to Option 40.

Upgrades

WCA2UP 02 - 65.5 MSample Deep Memory, Frequency Mask Trigger.

WCA2UP 03 - IQ, Differential IQ Inputs

WCA2UP 23 - W-CDMA Uplink Analysis (customer-installable).

WCA2UP 24 - GSM/EDGE Analysis (customer-installable).

WCA2UP 25 - cdma2000 1x Analysis (customer-installable).

WCA2UP 26 - 1xEV-DO Analysis (customer-installable).

WCA2UP 27 - 3GPP Release 5 Downlink (HSDPA) Analysis (customer-installable).

WCA2UP 28 - TD-SCDMA Analysis Software (customer-installable).

RSA34UP40 - 3GPP Release 6 (HSUPA) Analysis Software Upgrade (customer-installable)*8

WCA2UP 1A - External Preamp, 100 MHz — 3 GHz, 20 dB gain, 6.5 dB Noise Figure at 2 GHz (typical).

WCA2UP 1R - Rack mount kit upgrade for WCA200A Series.

WCA2UP IF - Installation labor for WCA2UPxx (no calibration required).

WCA2UP IFC - Installation labor for WCA2UPxx (with calibration).

RSA3SASW - USB Stand Alone Software Key.

*8 Option 23 and Option 27 are required for 3GPP Release 6 (HSUPA) Analysis in addition to Option 40 

RSA3303A, RSA3308A

Real-Time Spectrum Analyzer RSA3303A, DC – 3 GHz

Real-Time Spectrum Analyzer RSA3308A, DC – 8 GHz

Includes: User manual, Programmer's manual, Power cord, BNC-N adapter, USB Keyboard and Mouse.

Product Options*9

Opt. 02 - 65.5 MSample Deep Memory, Frequency Mask Trigger.

Opt. 03 - IQ, Differential IQ Inputs.

Opt. 21 - Advanced Measurement Suite Software.

Opt. 1A - External Preamp, 100 MHz – 3 GHz, 20 dB gain, 6.5 dB Noise Figure at 2 GHz (typical).

Opt. 1R - Rack mount kit.

Opt. SASW - USB Stand Alone Software Key.

Upgrades

RSA3UP 02 - 65.5 MSample Deep Memory, Frequency Mask Trigger.

RSA3UP 03 - IQ, Differential IQ Inputs.

RSA3UP 21 - Advanced Measurement Suite (customer-installable).

WCA2UP 1A - External Preamp, 100 MHz – 3 GHz, 20 dB gain, 6.5 dB Noise Figure at 2 GHz (typical).

WCA2UP 1R - Rack mount kit upgrade for RSA3300A Series.

RSA3UP IF - Installation labor for RSA3UPxx (no calibration required).

RSA3UP IFC - Installation labor for RSA3UPxx (with calibration).

RSA3SASW - USB Stand Alone Software Key.

*9 Specifications for Options 21 through 40 can be found in the Real-Time Spectrum Analyzer Software Options datasheet, on www.tektronix.com/RSA.

Accessories

RTPA2A - Probe Adapter box for TekConnect Probes.

119-4146-00 - RF Near Field Passive Probe Kit.

International Power Plugs

Opt. A0 - North America Power.

Opt. A1 - Universal EURO Power.

Opt. A2 - UK Power.

Opt. A3 - Australia Power.

Opt. A4 - North America Power, 240 V.

Opt. A5 - Switzerland Power.

Opt. A6 - Japan Power.

Opt. A10 - China Power.

Opt. A99 - No Power Cord.

Service Options

Opt. C3 - Calibration Service 3 Years.

Opt. C5 - Calibration Service 5 Years.

Opt. D1 - Calibration Data Report.

Opt. D3 - Calibration Data Report 3 Years (with Opt. C3).

Opt. D5 - Calibration Data Report 5 Years (with Opt. C5).

Opt. R3 - Repair Service 3 Years.

Opt. R5 - Repair Service 5 Years.

Language Options

Opt. L0 - English User/Programmer Manual.

Opt. L5 - Japanese User/Programmer Manual.