Question :
Can a User Script be loaded into 2600B series SMU and run from power on?
Réponses :
Yes, scripts can be loaded into the 2600B series SMU and run from power on using Test Script Builder.
- Write a script or function in Test Script Builder under a chosen project.
- In taskbar, select Run -> Run Configurations…
- Under TSP Script
- Select the drop-down arrow
- Select the project containing the desired scripts
- Under the Main Tab
- Select a TSP script from the Available Project Files
- Click Add >
- Scripts will be displayed in Load Order
- Under the Script Attributes Tab
- Select Load and Execute in Run Action Section
- Select Load and Debug in Debug Action
- Select Non-volatile and Auto Run in Storage
Click Run button in bottom right of window.
The above steps are applicable to any Keithley instrument that is compatible with Text Script Builder.