Settings saved on Model 6517B when updating firmware.

Yes, settings can be saved using the foll


This FAQ applies to:
6517B
468411
Differences between TCP202 and TCP202A

Differences are:

1. The jaw size of TCP202 is 3.8 mm while the jaw opening for TCP202A is 5 mm.

2. The TCP202A is rated for 150 V Cat on bare wire.


This FAQ applies to:
TCP202, TCP202A
468406
Size of TRCP current probe loops

The size of the loops are given in this document.


This FAQ applies to:
TRCP0300, TRCP0600, TRCP3000
468376
AFG3000B and AFG3000C differences

There at two differences.

1. The AFG3000C models the display was changed to TFT LCD which has a better viewing angle and brighter display.

2. And two new models were added. AFG3051C and AFG3052C.


This FAQ applies to:
AFG3011C, AFG3021B, AFG3021C, AFG3022B, AFG3022C, AFG3051C, AFG3052C, AFG3101C, AFG3102C, AFG3151C, AFG3152C, AFG3251C, AFG3252C
468371
Differences in the AFG1022 and AFG3022C

1. AFG1022 channel 2 is a continues signal only no modulation, burst, or sweep.  AFG3022C has full function on both channels.
2. AFG1022 Phasing of channel 1 and 2 is only available on for Sine wave below 1 MHz.  AFG3022C has phase control on any signal type and frequency.

This FAQ applies to:
AFG1022, AFG3022C

468361
How to Use a Link Partner in Testing with TDSET3 */
This FAQ applies to:
DPO4ENET, DPO5034B, DPO5054, DPO5054B, DPO5104, DPO5104B, DPO5204, DPO5204B, DPO70404, DPO70404B, DPO70404C, DPO7054, DPO70604, DPO70604B, DPO70604C, DPO70804, DPO70804B, DPO70804C, DPO7104, DPO71254, DPO71254B, DPO71254C, DPO71604, DPO71604B, DPO71604C, DPO72004, DPO72004B, DPO72004C, DPO72304DX, DPO72304SX, DPO72504DX, DPO7254, DPO73304DX, DPO73304SX, DPO7354, DPO75002SX, DSA70404, DSA70404B, DSA70404C, DSA70604, DSA70604B, DSA70604C, DSA70804, DSA70804B, DSA70804C, DSA71254, DSA71254B, DSA71254C, DSA71604, DSA71604B, DSA71604C, DSA72004, DSA72004B, DSA72004C, MSO5034, MSO5034B, MSO5054, MSO5054B, MSO5104, MSO5104B, MSO5204, MSO5204B, MSO70404, MSO70404C, MSO70604, MSO70604C, MSO70804, MSO70804C, MSO71254, MSO71254C, MSO71604, MSO71604C, MSO72004, MSO72004C, MSO72304DX, MSO72504DX, MSO73304DX, Opt. 5-SRENET, Opt. ET3, Opt. SR-ENET
468146
Could the number of analog channel be upgraded on the 5 series MSO?

No.  The analog channel count is fixed at the initial purchase.


This FAQ applies to:
MSO54, MSO56, MSO58
468131
What is the range of the logic threshold supported on the 5 and 6 Series MSO?

The logic threshold can be set between +/-40V with the TLP058 logic probe. Each individual digital channel can be set for its own threshold level.


This FAQ applies to:
MSO54, MSO56, MSO58, MSO64, TLP058
468126
How many digital channels are supported on the 5 and 6 Series MSO?

Each analog channel on the 5 and 6 Series MSO can be converted into 8 digital channels with the logic probe TLP058. 

For a 4-channel model (MSO54, MSO64), it is up to 32 digital channels.

For a 6-channel model (MSO56), it is up to 48 digital channels.


This FAQ applies to:
MSO54, MSO56, MSO58, MSO64
468121
PicoAmmeters/Electromeers/Sensitive SMUs -What is Guard? */ /*-->*/


This FAQ applies to:
6482, 6485, 6487, 6514, 6517B