Download

Download Manuals, Datasheets, Software and more:

DOWNLOAD TYPE
MODEL or KEYWORD

Feedback

What is the maximum number of samples that can be analyzed by TDSJIT3 Advanced/Essentials on the DPO7000 series scopes?

Question :

What is the maximum number of samples that can be analyzed by TDSJIT3 Advanced/Essentials on the DPO7000 series scopes?

Answer :

TDSJIT3 Advanced/Essentials V3.0.0.7 is limited to a testing a maximum of 64 M on a DPO7000 series scopes with longer waveform record length?  You can select the longer records while running TDSJIT3, but only the first 64 M will be analyzed by the application.


This FAQ Applies to:

No product series

Product:

FAQ ID 64181

View all FAQs »