Digital Storage Oscilloscope

TDS6000B/C Series

Features & Benefits

This product is in limited supply.
  • Bandwidths of 15 GHz (TDS6154C), 12 GHz (TDS6124C), 8 GHz (TDS6804B), and 6 GHz (TDS6604B)
  • Risetimes to 19 ps 20-80% (28 ps 10-90%) on TDS6154C typical risetime, with channel-matched, user-selectable DSP
  • 40 GS/s real-time sample rate on two channels*1, 20 GS/s real-time sample rate on all four channels with 500 fs/sample interpolated points
  • Up to 64 Megasamples on two channels*1, up to 32 Megasamples record length on all four channels with MultiView Zoom™ function for quick navigation
  • MyScope® Custom Control Windows Enhance Productivity
  • Right Mouse Click Menus for Exceptional Efficiency
  • Pinpoint™ triggering provides the most flexible and highest performance triggering, with over 1400 combinations to address virtually any triggering situation
  • Serial pattern triggering up to 3.125 Gb/s with 8b/10b protocol triggering
  • Serial data analysis and compliance at rates of 6.25 Gb/s and above
  • OpenChoice® software with Microsoft Windows XP OS delivers built-in networking and analysis
  • Technology specific software solutions provide built-in domain expertise for serial data, jitter, ethernet, DVI, USB2.0, communications, and power measurements
  • System includes: Dual processor system (2.8 GHz Pentium 4 and 583 MHz PowerPC), high resolution XGA display, front panel CD-R/W, front panel USB2.0 port, and 1000BaseT network connection

Applications

  • Signal Integrity, Jitter, and Timing Analysis
  • Validation, Debug, Characterization, and Compliance of Next Generation Digital Designs
  • Computer, Datacom, Storage Area Network Equipment Designs and High-speed Backplanes
  • High Energy Physics measurements and Data Acquisition

*1 For C Model versions only.

Uncompromised Performance Oscilloscope Solution — Probing, Acquisition, Analysis, Compliance, and Debug

TDS6000B/C Series digital storage oscilloscopes provide unprecedented performance along with a complete feature set designed to address design validation, debug, and compliance challenges of next generation computer, datacom, and communications equipment. High bandwidth, high sample rate, and deep memory also provide the ideal solution for data acquisition applications.

Uncompromised Acquisition

You won’t need to need to trade-off bandwidth, record length, and sample rate for your serial data measurement and analysis needs. The TDS6000C DSOs provides acquisition architecture with 40 GS/s maximum sample rate and 64 Megasamples record length on two channels (20 GS/s and 32 Megasamples on each of the four channels simultaneously), for the acquisition power you need. They provide the ultimate combination of bandwidth, sample rate, and record length for the fastest signals. The TDS6154C provides matched 15 GHz performance across any two channels using advanced, Tektronix-proprietary DSP enhancement, important for high speed channel-channel measurements. The user-selectable DSP filter on each channel provides magnitude and phase correction, plus extension of the analog bandwidth to 15 GHz for more accurate signal fidelity on high speed measurements — easily capturing the fifth harmonic of 3.0 GHz embedded clocks used in next generation 6.0 Gb/s serial data standards, and even the third harmonic of 5 GHz clocks being developed for future systems. The DSP filter on each channel can also be switched-off to take advantage of true 12 GHz analog bandwidth for applications needing the highest-available raw data capture.

MyScope® Custom Control Windows

MyScope control windows allow you to build your own control windows with only the controls, features, and capabilities that you care about and are important in your job. For the first time you can create your own personalized “toolbox” of oscilloscope features. No longer do you need to search through menus for features or re-learn how to drive the oscilloscope after a break from the lab. MyScope control windows are easily created in a matter of minutes using a simple, visual, drag and drop process. Once created, these customized windows are easily accessed through a dedicated MyScope button and menu selection on the oscilloscope button/menu bar, just like any other control window. You can make an unlimited number of custom control windows, enabling each person who uses the oscilloscope, in a shared environment, to have their own unique control window. Since the control windows are stored as files on the hard drive, they can easily be transferred to other TDS5000B or TDS/CSA7000B Series oscilloscopes, or they can even be e-mailed to a co-worker around the world when the need arises. MyScope control windows will benefit all oscilloscope users, from eliminating the ramp-up time that many face when returning to the lab after not using an oscilloscope for a while, to the power user who can now operate far more efficiently. Everything you need is found in one control window rather than having to constantly navigate through menu after menu to repeat similar tasks.

Right Clicks

Right mouse click menus make simple things as they should be – simple. Right click menus are context sensitive, meaning the choices presented in the menu depend on where you right clicked the mouse. This makes right click menus extremely intuitive. Want to change the cursor type? Right click on a cursor or the cursor readouts. Want to change the reference levels of an automatic measurement? Right click on the measurement. Want to change trigger parameters? Right click on the trigger readouts. Want to change a waveform’s color? Right click on the waveform handle. Virtually all objects on the oscilloscope display have right click menus associated with them that include all the appropriate actions or features relative to those objects. There are also right click menus for regions of the display in addition to just objects. For example, right clicking in the main graticule brings up a menu with choices such as Clear Data, Default Setup, Autoset, Screen Captures, Save All Waveforms, and Add Screen Text, providing single click access to many of your most commonly performed tasks.

Pinpoint™ Triggering

The ability to trigger an oscilloscope on events of interest is paramount in high speed debug and validation. Whether you're trying to find a system error, or need to isolate a section of a complex signal for further analysis, Tektronix’ Pinpoint triggering provides the solution. The Pinpoint trigger system uses Silicon Germanium (SiGe) technology to provide trigger sensitivity of up to 9 GHz (TDS6000C models), and allows selection of all trigger types on both A and B trigger circuits. It can capture glitches down to 100 ps wide with 1 psRMS trigger jitter typical (TDS6000C models). Other trigger systems offer multiple trigger types only on a single event (A Event), with delayed trigger (B Event) selection limited to edge type triggering and often does not provide a way to reset the trigger sequence if the B-event doesn't occur. But Pinpoint triggering provides the full suite of advance trigger types on both A and B triggers with Reset triggering to begin the trigger sequence again after a specified time, state, or transition so that even events in the most complex signals can be captured. Other oscilloscopes typically offer less than 20 trigger combinations; Pinpoint triggering offers over 1400 combinations, all at full performance.


Trigger on glitches down to 100 ps wide.


Pinpoint trigger system provides the most advanced, highest performance triggering available.

Protocol Triggering and Decoding Software (Opt. PTD)

Easily decode 8b/10b and other encoded serial data streams, and set desired encoded words for the serial pattern trigger to capture. 8b/10b decoding on data rates over 10 GB/s is possible on the TDS6000C models. Option PTD recovers the clock signal, identifies the transitions, decodes the characters and other protocol data. The TDS6000C models can trigger on up to four consecutive 10-bit words or specified error conditions at data rates up to 3.125 Gb/s. And you can see the captured bit sequences decoded into their words for convenient analysis.


Decode 8b/10b data streams and set up to 4 data words (40 bits) to trigger on.

Unparalleled Analysis

Waveform data analysis can take many forms. Whether it’s a simple math expression, waveform mask testing, a pass/fail compliance test, or a custom application that you develop, the TDS6000B/C Series offer the industry's most comprehensive set of analysis and compliance tools.

Built-in Analysis Tools

Standard tools built into the TDS6000B/C offer a wide range of analysis capabilities including Cursors, Measurements, Math Equation Editor, Serial Data Communications Mask Testing (with Opt. SM), and Spectrum Analysis.


Measurement System. Enables over 50 parametric measurements in the amplitude, time, and statistical domain.


8.5 Gb/s TDSRT-Eye™ diagram on the TDS6154C.


Waveform Math Equation Editor. Enables boundless analysis on waveform data.

Technology-specific Software Solutions

Validation, debug, and compliance testing often require automated test tools that quickly analyze waveform data and provide the answers you need. The TDS6000B/C models provide complete “turn-key” solutions for the most demanding technologies. Software options are also available for performing validation and compliance measurements on emerging industry standards.


TDSRT-Eye™ software. Compliance and analysis for testing high speed serial standards. Eye diagrams at data rates to beyond 6.25 Gb/s. Software clock recovery to ≥10 Gb/s.


TDSJIT3 v2.0 Jitter analysis for validation and debug high speed digital systems.


TDSET3 10/100/1000BaseT Compliance.


TDSDVI Digital Visual Interface Compliance.

OpenChoice® Analysis

Designing your own custom solution? The analysis and networking features of OpenChoice software adds more flexibility to Tektronix open Windows XP oscilloscopes:

  • • Fast, PCI bus speed communication between the data acquisition processor and the Microsoft Windows desktop.
  • • ActiveX controls to connect the oscilloscope to popular Windows applications - WITHOUT leaving the application.
  • • PnP drivers to control the scope from LabVIEW and Lab Windows/CVI running directly on the oscilloscope, or running on external PCs.

Access data seamlessly for your own custom application such as this Microsoft Excel example.

Support for application development environments includes Visual BASIC, .NET, C, C++, MATLAB, LabVIEW and LabWindows/CVI.

Differential Probing with Performance to >12 GHz

Most high speed signals today are differential. The P7313 12.5 GHz Differential Probing System provides true differential connection to the device under test for a variety of connection requirements: solder-in, hand held, and fixtured. Typical system performance with the TDS6154C exceeds 12.5 GHz bandwidth. The P7313 offers the lowest loading, highest signal fidelity, and lowest cost per connection in the industry. Versatile, inexpensive Tip-Clip™ adapters provide the optimum solution for virtually any connection need.

The P7380SMA Differential Probing System provides a 50 Ω per side termination network with a termination voltage. The termination voltage can be applied externally or through the TekConnect® interface from the TDS6000B/C probe menu, or not used at all. A gain switch provides two different sensitivity settings of the probe and an Aux Out provides an inverted version of the signal for driving other equipment. The P7380SMA probing system used with the TDS6000B/C oscilloscopes provides an ideal solution for validation and compliance testing of high speed serial data links.


P7313 12.5 GHz and P7380 8 GHz Differential Probes.

back to top
- Introduction
Characteristics/Specs
Ordering Information
 Print Data Sheet (289kB)
Check List Price width=10
width=1
Request a Quote width=10
width=1
 Waveform File Format (189kB) width=10
width=10

Product(s) are manufactured in ISO registered facilities.

55W-14873-9,  26-JUL-2006