Digital Phosphor Oscilloscopes

TDS3000C Series

Features & Benefits

Key Performance Specifications
  • 100 MHz, 300 MHz, and 500 MHz Bandwidth Models
  • 2 or 4 Channels
  • Sample Rates up to 5 GS/s Real Time on All Channels
  • 10 k Standard Record Length on All Channels
  • 3,600 wfms/s Continuous Waveform Capture Rate
  • Suite of Advanced Triggers
Ease-of-Use Features
  • Front-panel USB Host Port for Easy Storage and Transfer of Measurement Data
  • 25 Automatic Measurements
  • FFT Standard
  • Multiple Language User Interface
  • WaveAlert® Automatic Waveform Anomaly Detection
  • TekProbe® Interface Supports Active, Differential, and Current Probes for Automatic Scaling and Units
Portable Design
  • Lightweight Design (only 7 lb./3.2 kg) for Easy Transport
  • Optional Internal Battery Operation Provides up to Three Hours Without Line Power
Application Modules for Specialized Analysis
  • Advanced Analysis Module
  • Limit Testing Module
  • Telecommunications Mask Testing Module
  • Extended Video Module
  • 601 Serial Digital Video Module

Applications

  • Digital Design and Debug
  • Video Installation and Service
  • Power Supply Design
  • Education and Training
  • Telecommunications Mask Testing
  • Manufacturing Test
  • General Bench Testing

Performance You Need at a Price You Can Afford

Performance in an Affordable Package

The TDS3000C Series Digital Phosphor Oscilloscopes (DPO) provide you with the performance you need at a price you can afford. Bandwidths range from 100 MHz to 500 MHz, with up to 5 GS/s sample rates for accurate representation of your signal.

DPOs Provide a Greater Level of Insight into Complex Signals

To solve a problem, first you need to see it. The TDS3000C Series combines 3,600 wfms/s continuous waveform capture rate and real-time intensity grading so you can see the problem and solve it.

Fast waveform capture rates on a continuous basis save time by quickly revealing the nature of faults so advanced triggers can be applied to isolate them.

Real-time intensity grading highlights the details about the history of a signal’s activity, making it easier to understand the characteristics of the waveforms you’ve captured. Unlike other comparable oscilloscopes, the history remains even after the acquisition is stopped.

Quickly Debug and Characterize Signals with DRT Sampling Technology and sin(x)/x Interpolation

The TDS3000C Series combines unique digital real-time (DRT) sampling technology with sin(x)/x interpolation to allow you to accurately characterize a wide range of signal types on all channels simultaneously. With the TDS3000C Series there is no change in sampling rate when additional channels are turned on, unlike other comparable oscilloscopes. This sampling technology makes it possible to capture high-frequency information, such as glitches and edge anomalies, that elude other oscilloscopes in its class, while sin(x)/x interpolation ensures precise reconstruction of each waveform.


Look for unintentional circuit noise with the TDS3000C Series’ FFT capability.


Custom video triggering allows the TDS3000C Series to trigger on standards such as RS-343 (26.2 kHz scan rate).


The TDS3000C Series provides breakthrough test speeds for telecommunications line card testing. The telecom QUICKMENU puts all the commonly used telecom test functions on a single menu.


The TDS3000C Series with the TDS3LIM module is ideal for manufacturing test applications where fast Go/No-Go decisions are required.


Trace and identify ITU-R BT.601 video signals with the TDS3SDI 601 Serial Digital Video Module.


WaveAlert waveform anomaly detection alerts you to any waveform that deviates from the “normal” input such as the glitch on channel 2.


Easily transfer, document, and analyze data on your PC.

Easy Setup and Use

When working under tight deadlines, you need your oscilloscope to be intuitive; you want to minimize time spent learning and relearning how to use it. The TDS3000C Series oscilloscopes help reduce your learning curve. Simple navigation and dedicated front-panel controls get you to where you want to be quickly, so that you spend less time learning and more time on the task at hand.

Simple Documentation and Analysis

The TDS3000C Series comes equipped with a USB host port so you can easily store and transfer measurement information to your PC.

OpenChoice® PC Communication Software allows you to simply pull screen images and waveform data into a standalone desktop application or directly into Microsoft Word and Excel.

To complement OpenChoice, National Instruments LabVIEW SignalExpress™ Tektronix Edition Software provides you with extended capabilities including advanced analysis, data logging, remote instrument control, and live waveform analysis.

If you prefer not to use a PC for analysis, the TDS3000C Series comes standard with 25 automatic measurements, waveform add, subtract, divide, and multiply math functions, and Fast Fourier Transform (FFT). Unlike other comparable oscilloscopes, the TDS3000C Series math and measurement allows you to use the full acquisition record length or isolate a specific occurrence within an acquisition.

Instrument Control

Utilizing the built-in Ethernet port, e*Scope web-based remote control allows you to a control TDS3000C Series oscilloscope from anywhere, using the Internet and your PC.

Work Where you Need to

The TDS3000C Series packs the power of a DPO in a compact design that is only 5.9 in. (149 mm) deep, freeing up valuable benchtop space. And when you need to move your oscilloscope to another lab, its portable 7 lb. (3.2 kg) design makes for easy transport.

If your work demands even more mobility, then the optional battery pack will give you up to three hours of operation without line power.


TDS3BATC provides you with up to three hours of portable battery operation.

Characteristics

TDS3000C Series Electrical Characteristics

Characteristic

TDS3012C

TDS3014C

TDS3032C

TDS3034C

TDS3052C

TDS3054C

Bandwidth

100 MHz

100 MHz

300 MHz

300 MHz

500 MHz

500 MHz

Calculated Rise Time (typical)

3.5 ns

3.5 ns

1.2 ns

1.2 ns

0.7 ns

0.7 ns

Input Channels

2

4

2

4

2

4

External Trigger Input

Included on all models

Sample Rate on each channel

1.25 GS/s

1.25 GS/s

2.5 GS/s

2.5 GS/s

5 GS/s

5 GS/s

Record Length

10 kpoints

Vertical Resolution

9 bits

Vertical Sensitivity, 1 MΩ

1 mV/div to 10 V/div

Vertical Sensitivity, 50 Ω

1 mV/div to 1 V/div

Input Coupling

AC, DC, GND

Input Impedance

1 MΩ in parallel with 13 pF or 50 Ω

DC Gain Accuracy

±2%

Maximum Input Voltage, 1 MΩ

150 VRMS with peaks at ≤400 V

Maximum Input Voltage, 50 Ω

5 VRMS with peaks at ≤30 V

Position Range

±5 div

Bandwidth Limit

20 MHz

20 MHz

20 MHz, 150 MHz

20 MHz, 150 MHz

20 MHz, 150 MHz

20 MHz, 150 MHz

Time Base Range

4 ns to 10 s

4 ns to 10 s

2 ns to 10 s

2 ns to 10 s

1 ns to 10 s

1 ns to 10 s

Time Base Accuracy

±20 ppm over any 1 ms time interval

Input/Output Interfaces

Ethernet Port

RJ-45 connector, supports 10Base-T LAN

USB Port

Front-panel USB 2.0 host port

Supports USB flash drive

GPIB Port

Full talk/listen modes, setting and measurements

(Optional with TDS3GV Communications Module)

RS-232-C Port

DB-9 male connector, full talk/listen modes; control of all modes, settings and measurements

Baud rates up to 38,400

(Optional with TDS3GV Communications Module)

VGA Video Port

DB-15 female connector, monitor output for direct display on large VGA-equipped monitors

(Optional with TDS3GV Communications Module)

External Trigger Input

BNC connector, input impedance >1 MΩ in parallel with 17 pF; max input voltage is 150 VRMS

Acquisition Modes

Mode

Description

DPO

Captures and displays complex waveforms, random events and subtle patterns in actual signal behavior. DPOs provide 3 dimensions of signal information in real time: Amplitude, time, and the distribution of amplitude over time

Peak Detect

High-frequency and random glitch capture. Captures glitches as narrow as 1 ns (typical) using acquisition hardware at all time base settings

WaveAlert®

Monitors the incoming signals on all channels and alerts the user to any waveform that deviates from the normal waveform being acquired

Sample

Sample data only

Average

Waveform averaged, selectable from 2 to 512

Envelope

Min-max values acquired over one or more acquisitions

Single Sequence

Use the Single Sequence button to capture a single triggered acquisition sequence at a time

Trigger System

Characteristic

Description

Main Trigger Modes

Auto (supports Roll Mode for 40 ms/div and slower), Normal, Single Sequence

B Trigger

Trigger after time or events

Trigger After Time Range

13.2 ns to 50 s

Trigger After Events Range

1 to 9,999,999 events

Trigger Types

Trigger

Description

Edge

Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF Reject, LF Reject

Video

Trigger on all lines or individual lines, odd/even or all fields on NTSC, PAL, SECAM

Extended Video

Trigger on specific lines in broadcast and non-broadcast (custom) standards and on analog HDTV formats (1080i, 1080p, 720p, 480p). Requires TDS3VID or TDS3SDI application module

Pulse Width

(or Glitch)

Trigger on a pulse width <, >, =, ≠ to a selectable time limit ranging from 39.6 ns to 50 s

Runt

Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again

Slew Rate

Trigger on pulse edge rates that are either faster or slower than a set rate. Edges can be rising, falling, or either

Pattern

Specifies AND, OR, NAND, NOR when true or false for a specific time

State

Any logic state. Triggerable on rising or falling edge of a clock. Logic triggers can be used on combinations of 2 inputs (not 4)

Comm

Provides isolated pulse triggering required to perform DS1/DS3 telecommunications mask testing per ANSI T1.102 standard. Requires TDS3TMT application module

Alternate

Sequentially uses each active channel as a trigger source

Waveform Measurements

Characteristic

Description

Cursors

Amplitude, Time

Automatic Measurements

Display any four measurements from any combination of waveforms. Or display all measurements with measurement snapshot feature. Measurements include Period, Frequency, +Width, -Width, Rise time, Fall time, +Duty cycle, -Duty cycle, +Overshoot, High, Low, Max, Min, Peak-to-peak, Amplitude, Mean, Cycle mean, RMS, Cycle RMS, Burst width, Delay, Phase, Area*1, Cycle Area*1

Measurement Statistics

Mean, Min, Max, Standard deviation. Requires TDS3AAM application module

Thresholds

User-definable thresholds for automatic measurements; settable in percent or voltage

Gating

Isolate a specific occurrence within an acquisition to take measurements, using either the screen or cursors

*1 Requires TDS3AAM application module.

Waveform Math

Characteristic

Description

Arithmetic

Add, subtract, multiply, and divide waveforms

FFT

Spectral magnitude. Set FFT vertical scale to Linear RMS or dBV RMS, and FFT window to Rectangular, Hamming, Hanning, or Blackman-Harris

Advanced Math*1

Integrate, Differentiate, Define extensive algebraic expressions including analog waveforms, math functions, scalars, up to two user-adjustable variables and results of parametric measurements. For example: (Intg (Ch1-Mean(Ch1)) × 1.414 × VAR1)

Note: Requires TDS3AAM application module.

*1 Requires TDS3AAM application module.

Waveform Processing

Characteristic

Description

Autoset

Single-button, automatic setup of all channels for vertical, horizontal and trigger systems, with undo autoset

Deskew

Channel-to-channel deskew ±10 ns may be manually entered for better timing measurements and more accurate math waveforms

Display Characteristics

Characteristic

Description

Display Type

6.5 in. (165.1 mm) liquid-crystal TFT color display

Display Resolution

640 horizontal × 480 vertical pixels (VGA)

Interpolation

Sin(x)/x

Waveform styles

Dots, vectors, variable persistence, infinite persistence

Graticules

Full, grid, crosshair, and frame. NTSC, PAL, SECAM, and vectorscope (100% and 75% color bars) with optional TDS3VID or TDS3SDI application modules

Format

YT, XY, and Gated XYZ (XY with Z-axis blanking available on 4-channel instruments only)

Power Source

Characteristic

Description

AC line power

   

   Source voltage

100 VRMS to 240 VRMS ±10%

   Source frequency

45 Hz to 440 Hz from 100 V to 120 V

45 Hz to 66 Hz from 120 V to 240 V

   Power consumption

75 W maximum

Battery power

Requires TDS3BATC, rechargeable lithium ion battery pack

   Operating time, typical

3 hours

Environmental and Safety

Characteristic

Operating

Non-operating

Temperature

0 ºC to +50 ºC

-40 ºC to +71 ºC

Humidity

Operating and Nonoperating:

Up to 95% RH at or below +30 ºC

Operating and Nonoperating:

Up to 45% RH +30 ºC up to +50 ºC

Altitude

To 3,000 m

15,000 m

Electromagnetic Compatibility

Meets or exceeds EN61326 Class A, Annex D radiated and conducted emissions and immunity; EN6100-3-2 AC Powerline Harmonic Emissions; EN6100-3-3 Voltage Changes, Fluctuation, and Flicker; FCC 47 CFR, Part 15, Subpart B, Class A; Australian EMC framework

Safety

UL61010B-1, CSA1010.1, IEC61010-1, EN61010-1

Physical Characteristics

Instrument

Dimensions

mm

in.

Width

375.0

14.8

Height

176.0

6.9

Depth

149.0

5.9

Weight

kg

lb.

Instrument only

3.2

7.0

   with accessories

4.5

9.8

Instrument Shipping

Package Dimensions

mm

in.

Width

502.0

19.8

Height

375.0

14.8

Depth

369.0

14.5

Rackmount (RM3000)

Dimension

mm

in.

Width

484.0

19.0

Height

178.0

7.0

Depth

152.0

6.0

Ordering Information

TDS3000C Series Digital Phosphor Oscilloscopes

Model

Description

TDS3012C

100 MHz, 2 Channel, 1.25 GS/s

TDS3014C

100 MHz, 4 Channel, 1.25 GS/s

TDS3032C

300 MHz, 2 Channel, 2.5 GS/s

TDS3034C

300 MHz, 4 Channel, 2.5 GS/s

TDS3052C

500 MHz, 2 Channel, 5 GS/s

TDS3054C

500 MHz, 4 Channel, 5 GS/s

Standard Accessories

Accessory

Description

One P6139B, 500 MHz, 10x passive probe per channel

 

User Manual and Translated Front-panel Overlay

Please specify preferred language option - see chart below

Power Cord

Please specify plug option - see chart below

Accessory Tray

 

Protective Front Cover

 

OpenChoice® PC Communication Software

Enables fast and easy communication between Windows PC and the TDS3000C Series through LAN, GPIB, or RS-232. Transfer and save settings, waveforms, measurements, and screen images

NI LabVIEW SignalExpress™ Tektronix Edition LE

A fully interactive measurement software environment optimized for the TDS3000C Series. Enables you to acquire, generate, analyze, compare, import, and save measurement data and signals using an intuitive drag-and-drop user interface that does not require any programming. Standard TDS3000C Series support for acquiring, controlling, viewing, and exporting your live signal data is permanently available through the software. A 30-day trial period of the full version provides additional signal processing, advance analysis, mixed signal, sweeping, limit testing, and user-defined step capabilities. Order SIGEXPTE for permanent full version capability

Traceable Certificate of Calibration

NIM/NIST

Documentation CD

 

3 Year Warranty

Covering all labor and parts excluding probes and accessories

Options

Option

Description

Opt. A0

North America

Opt. A1

Universal Euro

Opt. A2

United Kingdom

Opt. A3

Australia

Opt. A5

Switzerland

Opt. A6

Japan

Opt. A10

China

Opt. A11

India

Opt. A99

No power cord

Language Options

Option

Description

Opt. L0

English

Opt. L1

French

Opt. L2

Italian

Opt. L3

German

Opt. L4

Spanish

Opt. L5

Japanese

Opt. L6

Portuguese

Opt. L7

Simplified Chinese

Opt. L8

Traditional Chinese

Opt. L9

Korean

Opt. L10

Russian

Opt. L99

No manual

Recommended Accessories

Accessory

Description

TDS3GV

GPIB, VGA, RS-232 interface

TDS3AAM

Advanced Analysis Module. Adds extended math capability, arbitrary math expressions, measurement statistics, and additional automated measurements

TDS3LIM

Limit Testing Module. Adds custom waveform limit testing capabilities

TDS3TMT

Telecom Mask Testing Module. Adds pass/fail compliance of ITU-T G.703 and ANSI T1.102 standards, custom mask testing, and more

TDS3VID

Extended Video Analysis Module. Adds video quickmenu, autoset, hold, line count trigger, video picture mode, vectorscope*2 mode, HDTV format trigger graticules, and more

TDS3SDI

Serial/Digital Video Module. Adds 601 serial digital video to analog video conversion, video picture, vectorscope*2, and analog HDTV triggering capabilities, and more

TDS3BATC

Lithium-ion battery pack for up to 3 hours continuous operation without line power

AC3000

Soft case for carrying instrument

HCTEK4321

Hard plastic case for carrying instrument

(requires AC3000)

RM3000

Rackmount kit

SIGEXPTE

NI LabVIEW SignalExpress™ Tektronix Edition Software full version

Service Manual

English only (071-2507-xx)

TNGTDS01

Extensive instructions and step-by-step lab exercises provide education about the operation of TDS3000C Series Oscilloscopes. Kit includes self-paced CD-ROM based manual and signal source board. Optional hardcopy manual available for order separately

*2 Vectorscope does not support composite video.

Recommended Probes

Probe

Description

P6243

1 GHz, ≤1 pF input C 10x active probe

P5205

1.3 kV, 100 MHz high-voltage differential probe

P5210

5.6 kV, 50 MHz high-voltage differential probe

P5100

2.5 kV, 100x high-voltage passive probe

TCP202

50 MHz, 15 A AC/DC current probe

TCP303*3

15 MHz, 150 A current probe

TCP305*3

50 MHz, 50 A current probe

TCP312*3

100 MHz, 30 A current probe

TCPA300

100 MHz probe amplifier

TCP404XL*4

2 MHz, 500 A current probe

TCPA400

50 MHz probe amplifier

ADA400A

100x, 10x, 1x, 0.1x high-gain differential amplifier

*3 Requires TCPA300 probe amplifier.

*4 Requires TCPA400 probe amplifier.

Service Options

Option

Description

Available at time of purchase

Opt. CA1

Provides a single calibration event or coverage for the designated calibration interval, whichever comes first

Opt. D1

Calibration Data Report

Opt. R5

Repair Service - 5 year

Opt. SILV400

Standard Warranty Extended to 5 Years

Available after purchase

TDS30xxC-CA1

Provides a single calibration event or coverage for the designated calibration interval, whichever comes first

TDS30xxC-R1PW

Repair service coverage 1 year post warranty

TDS30xxC-R2PW

Repair service coverage 2 years post warranty

TDS30xxC-R5DW

Repair service coverage 5 years (includes product warranty period); 5-year period starts at time of customer instrument purchase

Select Service

X

STAY INFORMED: Events & Seminars | Email Newsletter | Follow Tektronix on Twitter Follow Tektronix on Facebook Follow Tektronix on Google+ Follow Tektronix on Youtube

MEMBERSHIP: Create an account for exclusive membership privileges. Learn More | Login