Characteristics
|
Characteristic
|
Description
|
|
Bandwidth (probe only)
|
≥1.5 GHz
|
|
Attenuation (probe only)
|
10x ±2%
|
|
Rise Time (probe only)
|
<267 ps
|
|
Input Capacitance
|
<1 pF
|
|
Input Resistance
|
1 MΩ
|
|
Linear Dynamic Range
|
±8 V (16 Vp-p)
|
|
Input Offset Range
|
±10 V
|
|
Max Input Voltage (nondestruct)
|
±15 V (DC + peak AC)
|
|
Propagation Delay
|
5.3 ns
|
Physical Characteristics
|
Probe Head Size
|
mm
|
in.
|
|
Height
|
11.25
|
0.44
|
|
Width
|
5.1
|
0.20
|
|
Length
|
60.4
|
2.38
|
|
Other Dimensions
|
|
Cable Length
|
1300
|
51
|
|
Weight
|
kg
|
lb.
|
|
Net
|
0.45
|
1
|
Power Requirements
TAP1500 is powered
directly by oscilloscopes with the TekVPI probe interface.
Recommended Oscilloscopes
Oscilloscopes with the TekVPI™
probe interface.
Note: For best probe support,
download and install the latest version of the oscilloscope software
from www.tektronix.com
Environmental
|
Characteristic
|
Description
|
|
Temperature
|
|
Operating
|
0 °C to +50 °C
|
|
Nonoperating
|
-40 °C to +71 °C
|
|
Humidity
|
|
Operating
|
5% to 95% Relative Humidity (RH) at up to +30 °C;
5% to 85% RH above +30 °C up to +50 °C; noncondensing
|
|
Nonoperating
|
5% to 95% Relative Humidity (RH) at up to +30 °C;
5% to 85% RH above +30 °C up to +75 °C; noncondensing
|
|
Altitude
|
|
Operating
|
Up to 3,000 m (10,000 ft.)
|
|
Nonoperating
|
Up to 15,240 m (50,000 ft.)
|
|
Regulatory
|
|
Compliance Labeling
|
WEEE (European Union)
|
Standard Warranty
One year parts
and labor.