Characteristics
|
Mode
|
Amplitude
|
Frequency
|
Rise Time
|
|
(all values
are Typical)
|
|
Mode 1
|
1 Vp-p (-1 V..0 V)
|
1 kHz
|
0.8 ns
|
|
Mode 2
|
2 Vp-p (-2 V..0 V)
|
1 kHz
|
1.6 ns
|
|
Mode 3
|
2 Vp-p (-1 V..+1 V)
|
1 MHz
|
2 ns
|
|
Mode 4
|
8 Vp-p (-4 V..+4 V)
|
1 MHz
|
4 ns
|
Terminates Into - 50 Ω
Expected Load - 50 Ω
Status LED Values
|
Status LED Color
|
Description
|
|
Amber
|
The generator is powered and the output at the BNC
cable is not enabled. This is the default status at power-on
|
|
Green
|
The generator is powered and the output at the BNC
cable is enabled
|
|
Red
|
A fault condition may have occurred. Disconnect
and reconnect the generator to clear the fault condition
|
Recommended Output Mode
|
Probe*1
|
Operation
|
TEK-DPG
Output Mode*2
|
067-1686-xx
Input
Port
|
|
The recommended output mode is determined
by the combination of oscilloscope model, voltage probe, and current
probe, according to the following table. See the user instructions
for the Power Measurement Deskew Fixture, Tektronix part number 067-1686-xx,
for deskew fixture operation.
|
|
Passive Voltage Probes
|
|
P6139A, P5050,
and Other 10x Passive
Probes
|
Compensate timing with respect to current probes
|
DPO3000: Mode 2
DPO/MSO4000: Mode
1
|
A or B, depending on current probe type
|
|
Active Voltage Probes
|
|
TAP1500, TAP2500, TAP3500
TDP0500,
TDP1000, TDP1500
P6243, P6245, P6246, P6247, P6248,
P6250, P6251
|
Compensate timing with respect to current probes
|
DPO3000: Mode 2
DPO/MSO4000: Mode
1
|
A or B, depending on current probe type
|
|
High-voltage Probes
|
|
P5200, P5205, P5210
P6135A
|
Compensate timing with respect to current probes
|
DPO3000: Mode 2 or 4*3
DPO/MSO4000 : Mode 1 or 4*3
|
A or B, depending
on current probe
type
|
|
Small Current Probes
|
|
TCP0030
TCP202, TCP312, TCP305
A6302, A6312
|
Compensate timing with respect to voltage probes
|
DPO3000: Mode 2 or 4*3
DPO/MSO4000: Mode 1 or 4*3
|
A
|
|
Large Current Probes
|
|
TCP0150
TCP303
A6303
|
Compensate timing with respect to voltage probes
|
DPO3000: Mode 4
DPO/MSO4000: Mode
4
|
B
|
*1 Note that some probe models listed
do not utilize the TekVPI interface. Probes which use the TekProbe-BNC
interface require the TPA-BNC adapter to connect to a TekVPI oscilloscope.
*2 TEK-DPG is not supported by the DPO7000 Series.
The Probe Calibration output from the DPO7000 Series generates the
appropriate source signal for the Power Measurement Deskew Fixture.
*3 Mode 4 is a universal mode for high attenuation
probes. High-voltage probes and current probes can always use Mode
4.
Physical Characteristics
|
Dimensions
|
cm
|
in.
|
|
Height
|
3.048
|
1.20
|
|
Width
|
4.064
|
1.60
|
|
Length
|
11.73
|
4.62
|
|
Cable Length
|
41.225
|
16.2
|
|
Weight
|
kg
|
lb.
|
|
Shipping
|
0.325
|
0.7175
|
Power Requirements
- TEK-DPG is powered directly from DPO3000, DPO/MSO4000, and
DPO7000 Series Oscilloscopes, using the TekVPI probe interface.
Environmental
|
Characteristic
|
Description
|
|
Temperature
|
|
Operating
|
0 °C to +50 °C
|
|
Nonoperating
|
-40 °C to +71 °C
|
|
Humidity
|
|
Operating
|
5% to 90% relative humidity (% RH) at up to +40 °C
5% to 60% RH above +40 °C up to +50 °C, noncondensing
|
|
Nonoperating
|
5% to 90% RH (relative humidity) at up to +40 °C
5% to 60% RH above +40 °C up to +50 °C, noncondensing
|
|
Altitude
|
|
Operating
|
Up to 3000 meters (9,843 feet)
|
|
Nonoperating
|
Up to 12,000 meters (39,370 feet)
|
|
Regulatory
|
|
Emissions Compliance
|
EN 55011, Class A
|
|
EMC Compliance
|
EN61326, Class A
|
|
Compliance Labeling
|
WEEE (European Union)
|
Standard Warranty
One-year parts and labor.
Recommended Oscilloscopes
- DPO3000 and DPO/MSO4000 Series Oscilloscopes
with TekVPI probe interface.
- TEK-DPG is not supported
by the DPO7000 Series Oscilloscopes, as these models provide the appropriate
source signal at the Probe Calibration output on the front-panel BNC
connector.